ECG-9020P心电图机在高压舱中的应用研究

Application of ECG-9020P Electrocardiograph in Hyperbaric Chamber
下载PDF
导出
摘要 目的:探讨心电图机在高压舱中应用的可行性。方法:使用ECG-9020P心电图机,将导联线及心电传感器放置在高压舱内,主机部分及电源线放置在舱外,利用高压舱舱体预留的穿舱孔,对其进行简单改造。选取9名受试者,分别在0.40、0.28、0.22、0.19、0.16和0.13 MPa压力环境下进行心电图检查。结果:在0.40 MPa压力环境下进行心电图检查9人次;在0.28、0.22、0.19、0.16和0.13 MPa压力环境下各进行心电图检查4人次。29份心电图导联连接正确,基线稳定,无明显干扰波,心电波形质量高,便于临床医生分析解读。结论:经过改造的ECG-9020P心电图机可在不同压力环境下正常进行心电图检查,为高压舱内诊断心脏功能异常提供了便利。 Objective To study the application of electrocardiograph in hyperbaric chamber. Methods ECG-9020P electrocar- diograph was reformed simply, in order that the electrocardiogram lead could be connected with the subjects through the ) pass-through of hyperbaric chamber. The electrocardiogram lead and transducer were put inside the hyperbaric chamber, and the main unit and power line were outside the hyperbaric chamber. Nine subjects went through electrocardiogram examination in hyperbaric chamber at different atmospheric pressures including 0.40, 0.28, 0.22, 0.19, 0.16 and 0.13 MPa. Results Nine sub- jects went through electrocardiogram examination at the atmospheric pressure of 0.40 MPa, and four subjects through electrocar- diogram examination at different atmospheric pressures including 0.28, 0.22, 0.19, 0.16 and 0.13 MPa. All of the 29 electrocar- diogram waveforms were in the normal range without any interference. The base lines were stable, and easily analyzed by doctors. Conclusion The reformed ECG-9020P electrocardiograph can work normally at different atmospheric pressures, which is conve- nient for doctors to diagnose cardiac functions. [Chinese Medical Equipment Journal, 2014,35 (2): 131-132,134]
出处 《医疗卫生装备》 CAS 2014年第2期131-132,134,共3页 Chinese Medical Equipment Journal
关键词 高压舱 高压氧 心电图机 hyperbaric chamber hyperbaric oxygen electrocardiograph
  • 相关文献

参考文献7

二级参考文献44

共引文献36

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部