摘要
目的比较Lenstar与IOL—Master及A超在测量眼轴长度(axial length,AL)、前房深度(anterior chamber depth,ACD)时的差异,分析三种测量方法的一致性,为Lenstar的临床应用提供依据。方法前瞻性临床研究。对2011年12月至2012年3月在北医三院眼科的白内障术前患者及志愿者计78例123只眼,分别应用Lenstar与IOL—Master及A超进行AL、ACD测量,Lenstar与IOL—Master及A超测量值间的差异采用配对t检验,相关性采用Pearson相关分析,一致性采用Bland—Ahman统计分析。结果Lenstar与IOL—Master测量AL及ACD的平均值差值分别为(-0.003±0.048)mm及(-0.044±0.091)mm,其差异均无统计学意义(P=0.517;P=0.573,P〉0.05)。Lentar与A超测量AL及ACD的平均值差值分别为(0.117±0.195)(0.080±0.253)mm,其差异均有统计学意义(P〈0.01)。三种测量方法所获得的AL及ACD均具有密切的线性相关差异有统计学意义(r≥0.916,P〈O.01)。Bland—Altman分析显示,在测量AL及ACD时,Lenstar与IOL—Master的一致性较好,与传统A超测量值的一致性较差。结论在测量眼轴长度及前房深度时,Lenstar与IOL—Master的一致性较好,与传统A超测量值的一致性较差,由于Lenstar具有一次测量可获得多项数据、简单快捷、非接触等优点,是一种在临床中可选择的准确性较高地生物学测量仪器。
Objective To evaluate a new noncomact optical biometer (Lenstar) using optical low-coherence reflectometry and to compare the biometric measurements (axial length, AL and anteri- or chamber depth, ACD) with those obtained from current clinical instrumentation (IOL-Master and ultrasound A-scan). Methods In this prospective clinical study, measurements of AL and ACD ob- tained from 123 eyes of 78 subjects with the Lenstar were compared with those obtained with the IOL-Master or ultrasound A-scan. The results were evaluated using Bland-Altman analyses. The differ- ences between both methods were assessed using the paired t test, and its correlation was evaluated by Pearson coefficient. Results Lenstar measurements of AL and ACD were similar to the IOL-Mas- ter (average difference=-0.003±0.048mm, P =0.517; average difference=-0.044±0.091mm, P =0.573), re- spectively. Lenstar measurements of AL and ACD were significantly longer than those of A-scan (av- erage difference=0.117±0.195mm, P 〈0.001; average difference=0.080±0.253mm, P 〈0.001), respective- ly. A Pearson correlation revealed a high correlation for AL and ACD (r ≥0.916 for all, P 〈0.001 for all). Bland-Altman plots showed good agreements for AL measurements and ACD between Len- star and IOL-Master and the level of agreement was not acceptable between Lenstar and A-scan. Conclusions For the measurement of AL and ACD, there is good agreement between Lenstar and IOL-Master. Nevertheless, Lenstar and A-scan cannot be used interchangeably. The results, coupled with a unique ability to analyze the ocular structures non-contactly make the Lenstar a promising new instrument for ocular evaluation in research and clinical practice,
出处
《中国实用眼科杂志》
CSCD
北大核心
2014年第3期314-318,共5页
Chinese Journal of Practical Ophthalmology