iDose^4双模型迭代重建算法对CT图像质量及扫描剂量影响的模体研究 被引量:11

Investigation impact of dual model iterative reconstruction method(iDose^4) on the CT image quality and dose reduction—a phantom approach
下载PDF
导出
摘要 目的:评价双模型迭代重建算法(iDose4)和滤波反投影算法(FBP)对CT图像质量的影响,探讨应用iDose4降低扫描剂量的潜力。方法:应用256层螺旋CT(Philips Brilliance iCT),在120 kV管电压条件下分别采用200/180/160/140/120/100/80/60/40/20 mAs条件下扫描Catphan模体,采用FBP和iDose4 1~6重建横断图像,以200 mAs FBP重建图像作为参照,分析扫描条件、重建算法对图像质量的影响(CT值、图像噪声、信噪比以及低对比度分辨力),并分析图像质量与扫描剂量的关系。结果:iDose4与FBP重建图像CT值有高度的一致性(P>0.05);iDose4级别1~6重建可以不同程度的降低图像噪声(9.46%~43.30%);在管电流降低至初始值40%~50%时(80~100 mAs),iDose4级别6重建图像与200 mAs FBP重建图像其图像噪声无统计学差异(P>0.05);管电流降低至初始值50%~60%情况下(100~120 mAs),其低对比度分辨力基本维持不变。结论:iDose4可以显著降低图像噪声,提高信噪比,改善图像质量。相对于FBP重建图像,利用iDose4高级迭代重建技术可以在扫描剂量隆低至40%~50%情况下,维持图像质量不变。 Objective: To investigate the impact of iterative reconstruction method (iDose4) on the CT image quality and dose reduction by using phantom studies. Methods: A Catphan phantom was scanned by a 256-slice CT (Philips Brilliance iCT) with varied mAs from 20 to 200, the axial images were reconstructed by filtered back projection (FBP) and iDose4 with a level of 1 to 6. Image quality was assessed with following aspect: CT value accuracy and consistency, image noise and sig- nal-noise-ratio (SNR), low density resolution. Meanwhile the volume CT dose index CTDIvol and Dose length product DLP were recorded for dose evaluation. Results: The axial image CT values has a high degree of consistency (P〉0.05) for both iDose4 and FBP reconstruction method. The image noise can be reduced in different degree with different iDose4 level 1-6 reconstruction (9.46%-43.30%). When tube current reduced to 40%-50%(80-100 mAs), iDose4 level 6 got image without significant difference regarding to the image noise compared to 200 mAs FBP image(P〉0.05); under the conditions of 50%-60% initial tube current (100-120 mAs), low density resolution could remain unchanged. Conclusion: iDose4 can significantly reduce image noise and improve image quality, under the condition of 40%-50% scanning dose can get satisfactory image quality. Thus iterative reconstruction has larger space to lower the scanning dose.
出处 《中国临床医学影像杂志》 CAS 2014年第3期190-194,共5页 Journal of China Clinic Medical Imaging
关键词 辐射剂量 图像重建 计算机辅助 体层摄影术 螺旋计算机 Radiation dosage Image processing, computer-assisted Tomography, spiral computed
  • 相关文献

参考文献15

  • 1Hausleiter J, Meyer T. Estimated radiation dose associated with cardiac CT angiography[J]. JAMA, 2009, 301(5): 500-507.
  • 2Brenner OJ, Hall EJ. Computed tomography-an increasing source of radiation exposure [J]. N Engl J Med, 2007, 357 (22): 2277- 2284.
  • 3Raman SP. CT Dose Reduction Applications Available Tools on the Latest Generation of CT Scanners[J]. J Am Col Radiol, 2013, 10(1): 37-4l.
  • 4Singh S, Kalra MK, Shenoy-Bhangle AS, et al. Radiation dose reduction with hybrid iterative reconstruction for pediatric CT[J]. Radiology, 2012, 263(2): 537-546.
  • 5Richard S, Husarik DB. Towards task -based assessment of CT performance: system and object MTF across different reconstruction algorithms[J]. Med Phys, 2012, 39(7): 4115-4122.
  • 6郭洪涛,彭明辰.医用诊断螺旋计算机断层摄影装置(CT)X射线辐射源检定规程[M].北京:中国计量出版社,2007:101-109.
  • 7Payne JT. CT radiation dose and image quality [J]. Radiol Clin North Am, 2005, 43(6): 953-962.
  • 8Lee CH, Goo JM. Radiation dose modulation techniques in the multi detector CT era: from basics to practice [J]. Radiographies, 2008, 28(5): 1451-1459.
  • 9Pickhardt PJ. Abdominal CT with model-based iterative reconstruction (MBIR): initial results of a prospective trial comparing ultralow-dose with standard-dose imaging [J]. AJR, 2012, 199(6): 1266-1274.
  • 10Hara AK. Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study [J]. AJR, 2009, 193 (3): 764-771.

二级参考文献8

  • 1John MB, Estella MG. Dose reduction in pediatric CT: a ra- tional approach [J]. Raiology, 2003, 228: 353-360.
  • 2De Maertelaer TD, Gevenois PA. VDose reduction in mul- tidetector CT using attenuation-based online tube current modulation [J]. AJR, 2003, 181: 331-334.
  • 3Mulkens TH, Bellinck P, Baeyaert M, et al. Use of an auto- matic exposure control mechanism for dose optimization in multidetector row CT examinations: clinical evaluation [J]. Radiology, 2005, 237: 213-223.
  • 4Kalra MK, Maher MM, Toth TL, et al. Strategies for CT radiation dose optimization [J].Radiology, 2004, 230:619 628.
  • 5Pontana F, Pagniez J, Flohr T, et al. Chest computed tomo graphy using iterative reconstruction vs filtered back projec tion (Part 1): evaluation of image noise reduction in 32 pa tient[J]. Eur RadioI, 2010, 21: 627-635.
  • 6Pontana F, Duhamel A, Pagniez J, et al. Chest computed tomography using iterative reconstruction vs filtered back projection (Part 2) : image quality of low-dose CT examina- tions in 80patients [J]. Eur Radiol, 2010, 21: 636-643.
  • 7Antonio Moscariello, Richard A P Takx , U. Joseph Seho- epf, Matthias Renker, et al. Coronary CT angiography: im age quality, diagnostic accuracy, and potential for radiation dose reduction using a novel iterative image reconstruction technique-comparison with traditional filtered back projection [J]. EurRadiol, 2011, 21: 2130-2138.
  • 8Anna Winklehner & Christoph Karlo, Gilbert Puippe, Bern- hard Schmidt, et al. Raw data-based iterative reconstruction in body CTA: evaluation of radiation dose saving potential [J]. EurRadiol, 2011, 11.

共引文献28

同被引文献88

引证文献11

二级引证文献33

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部