摘要 研制了一种铝靶软X射线源,这种X射线源对轻介质、薄样品的成像更清晰、分辨能力更强。利用多层滤片和生物样品对铝靶软X射线源进行了性能检测,实验显示铝靶软X射线源对轻介质和生物成像具有较高的吸收衬度,成像效果优于其他靶材X射线源,并通过改变工作电压可以显著调节谱线中软X射线的比例。这种铝靶X射线源适合作为实验室便携式软X射线光源。 An aluminum target soft X ray tube is presented. This soft X ray tube shows good performance on imaging thin or light material samples such as biological specimen or membranes with better resolution and definition. The experimental test of the soft X ray tube using multilayer membrane and biological samples is presented. The result shows that the imaging quality is better than ordinary X ray tube because more soft X ray is absorbed by the sample and the proportion of the soft X-ray in the emission spectrum is changed by adjust the work voltage. The aluminum target X ray tube is very useful in biological imaging, material inspection and X ray microscopy.
出处 《光学技术》 CAS CSCD 北大核心 2014年第2期184-188,共5页 Optical Technique
基金 国家自然科学基金项目(11105129)
关键词 X射线管 软X射线源 铝靶 X射线成像 X ray tube soft X ray source aluminum target X ray imaging
  • 相关文献

参考文献10

二级参考文献61

  • 1杨家敏,易荣清,马洪良,陈正林,王红斌,孙可煦,温树槐,郑志坚,唐道源,崔明启,黎刚,崔聪悟.透射光栅衍射率的实验标定和数值计算[J].光学学报,1996,16(11):1631-1635. 被引量:11
  • 2[1]Rayment T, Schroeder S M, Moggridge G D,et al. Electron-yield X-ray absorption spectroscopy with gas microstrip detectors.Rev Sci Instrum,2000,71(10):3640~3645
  • 3[2]Pogany A, Gao D, Silkins S W.Contrast and resolution in imaging with a microfocus X-ray source.Rev Sci Instrum,1997,68(7):2774~2782
  • 4[3]Stasio G D, Gilbert B, Nelson T,et al.Feasibility tests of transmission X-ray photoelectron emission microscopy of wet samples.Rev Sci Instrum,2000,71(1):11~14
  • 5[4]Wang K G, Niu H B.Study of the X-ray source with micro-beam. Proc SPIE,2001,4224:101~104
  • 6[5]Dhez P, Chevallier P, Lucatorto T B,et al.Instrumental aspects of X-ray microbeams in the range above 1 keV.Rev Sci Instrum, 1999,70(4):1907~1920
  • 7[6]Khacef A, Viladrosa R, Cachoncinlle C,et al.High repetition rate compact source of nanosecond pulses of 5~100 keV X-ray photos.Rev Sci Instrum,1997,68(6):2292~2297
  • 8[7]Ratliff L P, Bell E W, Parks D C, et al.Continuous highly charged ion beams from the National Institute of Stardards and technology electron-beam ion trap.Rev Sci Instrum,1997,68(5): 1998~2002
  • 9[8]Marrs R F, Schneider D H, MaDonald J W. Projection X-ray microscope powered by highly charged ions.Rev Sci Instrm,1998,69(1): 204~209