摘要 为缓解患者长时间使用透射式无创血氧饱和度监测仪带来的痛苦,提出采用反射式光电传感器进行脉搏血氧监测方法。利用LabVIEW软件强大的数据处理功能,结合NI USB-6353DAQ数据采集器,可快速实现系统的搭建。对10名患者进行脉率与血氧饱和度的监护测试实验,并与临床用血氧饱和度监测仪测量的数据对比,两者测量的血氧饱和度平均误差为0.2%,测量的脉率平均误差为0.8time/min。结果表明,本文设计与临床用血氧饱和度监测仪具有良好的一致性,能够进行脉率与血氧饱和度的测量,可以成为现有生理参数监测方法的一种补充。 In order to alleviate patients pain for the long time using transmission type non-invasive pulse and blood oxygen saturation monitor,a pulse and blood oxygen saturation monitoring method is proposed by using the reflection type photoelectric sensor. Using LabVIEW software with strong function of data processing, combined with the NI USB-6353 DAQ the data acquisition system can be built rapidly. Moni- toring test of pulse rate and oxygen saturation in 10 patients,and compared with the data of monitor and prototype, the oxygen saturation measured average error is 0.2 %0, and the average error of measurement of pulse rate is 0. 8 time/min. The results show that the design and clinical use of the blood oxygen satu ration monitor have good consistency, so this method can measure the pulse rate and blood oxygen satu- ration, can be a supplement to the existing physiological parameter monitoring method.
出处 《光电子.激光》 EI CAS CSCD 北大核心 2014年第3期620-624,共5页 Journal of Optoelectronics·Laser
关键词 反射式 传感 血氧饱和度 脉率 reflective sensor blood oxygen saturation pulse rate
  • 相关文献

参考文献16

  • 1HU Hong-jiao. Before and after abdominal operation in elderly patients with blood oxygen saturation monitoringsignificance[J]. Anhui Medical and Pharmaceutical Jour- nal,2010,4:25-29.
  • 2YIN Guang-zhi, XiNG Da, YANG Si-hua. Dynamic monito- ring of blood oxygen saturation in vivo using double-ring photoacoustic sensor [J]. Journal of Applied Physics, 2009,106(1) : 1186-1193.
  • 3Cicco G. A non invasive optical oximetry in humans., pre- liminary data[J]. Clinical hemorheology and microcircula- tion, 1999,21(3) : 311-314.
  • 4肖君,王龙,骆清铭,龚辉.基于近红外三波长的血糖检测系统研制[J].光电子.激光,2007,18(9):1135-1138. 被引量:10
  • 5Reekers M,Simon M J G,Boer F, et al. Pulse dye densi- tometry and indocyanine green plasma disappearance in ASA physical status I-II patients[J]. Anesthesia & Anal- gesia, 2010,110(2) : 466-472.
  • 6XlE Zheng-yang, HU Dan-feng, WANG Jia-jun. Develop- ment of non-invasive measurement and wireless monito- ring system of oxygen saturation[J]. Chinese Instruments, 2009,4:65-67.
  • 7胡静,王成,李章俊,马俊领.基于光电脉搏波描记方法的多生理参数测量研究[J].光电子.激光,2012,23(8):1631-1634. 被引量:15
  • 8Castillejo P, Martinez J F, Rodriguez-Molina J, et al. Inte- gration of wearable devices in a wireless sensor network for an E-health application[J]. Wireless Communications, IEEE, 2013,20(4) : 38-49.
  • 9SUN Xin, LIU Chang-chun, ZHAO Yu-juan, et al. Ankle bra- chial index measurement of photoelectric pulse sensor [J]. Journal of Optoelectronic · Laser, 2010, 21 (8): 1214-1217.
  • 10朱拓,陈国庆,朱益清,孔艳,薄亚明,倪晓武.长方体内多量散射介质光子密度波扩散方程研究[J].光电子.激光,2004,15(9):1132-1136. 被引量:2