摘要
Background Primary total knee arthroplasty(TKA) in severe valgus knees may prove challenging, and choice of implant depends on the severity of the valgus deformity and the extent of soft-tissue release. The purpose of this study was to review 8 to 11 years(mean, 10 years) follow-up results of primary TKA for varient-III valgus knee deformity with use of different type implants. Methods Between January 2002 and January 2005, 20 women and 12 men, aged 47 to 63(mean, 57.19±6.08) years old, with varient-III valgus knees underwent primary TKA. Of the 32 patients, 37 knees had varient-III deformities. Pie crusting was carefully performed with small, multiple inside-out incisions, bone resection balanced the knee in lieu of soft tissue releases that were not used in the series. Cruciate-retaining knees(Gemini MKII, Link Company, Germany) were used in 13 knees, Genesis II(Simth & Nephew Company, USA) in 14 knees, and hinged knee(Endo-Model Company, Germany) in 10 knees. In five patients with bilateral variant-III TKAs, three patients underwent 1-stage bilateral procedures, and two underwent 2-stage procedures. All implants were cemented and the patella was not resurfaced. The Hospital for Special Surgery(HSS) knee score was assessed. Patients were followed up from 8 to 11 years.Results The mean HSS knee score were improved from 50.33±11.60 to 90.06±3.07(P <0.001). The mean tibiofemoral alignment were improved from valgus 32.72o±9.68 o pre-operation to 4.89o±0.90 o post-operation(P <0.001). The mean range of motion were improved from 93.72o±23.69 o pre-operation to 116.61±16.29 o post-operation(P <0.001). No patients underwent revision. One patient underwent open reduction and internal fixation using femoral condylar plates for supracondylar femoral fractures secondary to a fall at three years. Three patients developed transient peroneal nerve palsies, which resolved within nine months. Two patients developed symptomatic deep vein thrombosis that was managed with rivaroxaban and thrombo-embolic deterrent stockings. There was no incidence of pulmonary embolism. Postoperative patient satisfaction was 80.7±10.4 points in the groups. Prosthetic survival rate was 100% at mean 10 years postoperative.Conclusions Not only hinged implants can be successfully used in variant-III valgus knees. As our results show, if proper ligament balancing techniques are used and proper ligament balance is attained, the knee may not require the use of a more constrained components. Our results also present alternative implant choices for severe knee deformities.