Sysmex XE-2100全自动血液分析仪的性能验证 被引量:3

Performance Verification of Sysmex XE-2100 Automated Hematology Analyzer
下载PDF
导出
摘要 目的:通过对Sysmex XE-2100全自动血液分析仪进行性能验证,了解和掌握该仪器的精密度、正确度、线性、携带污染率、功能灵敏度、可稀释范围、参考区间及手动与自动进样模式结果比对的性能。方法:以国际血液学标准化委员会、临床和实验室标准协会制定的评价方案为标准进行评价试验。结果:全血细胞计数精密度、正确度、线性良好,携带污染率低,功能灵敏度符合要求,可稀释范围回收率良好,参考区间适用。2种进样模式相关性好,与手工法白细胞分类结果比较,中性粒细胞、淋巴细胞、嗜酸性细胞的γ值均达0.9以上。异常细胞检测能力总有效性为95.4%。结论:Sysmex XE-2100全自动血液细胞分析仪各方面性能良好、测定准确可靠,可快速、有效地完成大批量标本的筛检。 Objective To learn the precision, correctness, linearity, residual contamination rate, function sensitivity, dilution range, reference interval, and manual or automatic sampling mode of Sysmex XE-2100 Automated Hematology Analyzer by per- formance verification. Mecdaods The verification was performed on the bases of the evaluation procedures from International Com- mittee for Standardization in Hematology and American Clinical Laboratory Standards Institute. ~ When used for complete blood count, the analyzer behaved well in precision, correctness, linearity, residual contamination avoidance, function sensitivity, dilution range and reference interval, and there was a high correlation between the two sampling modes; when used for manual leukocyte differential count, the analyzer had T values of neutrophilic granulocyte, achroacyte and eosinophile granulocyte more than 0.9, and the total effective rate for abnormal cell detection 95.4%. Conclusion Sysmex XE-2100 Automated Hematology Analyzer is effective, accurate and reliable, and can be used for mass samples screening.
出处 《医疗卫生装备》 CAS 2014年第3期85-88,共4页 Chinese Medical Equipment Journal
关键词 全自动血液分析仪 性能验证 精密度 携带污染率 automated hematology analyzer performance verification precision residual contamination rate
  • 相关文献

参考文献9

  • 1ISO15189医学实验室质量和能力认可准则[S].北京:中国合格评定国家认可委员会.2008:1236.
  • 2Hedley B D, Keeney M, Chin-Yee I,et d. Initial performance evalu- ation of the UniCel(R) DxH 800 Coulter(R) cellular analysis system[J]. Int J Lab Hematol, 2011,33 ( 1 ) :45-56.
  • 3叶应妩.全国I艋床检验操作规程[M].3版.南京:东南大学出版社.2006:496.
  • 4CLSI. EP52A Evaluation of precision performance of clinical chem- istry devices[S].
  • 5Meintker L, Ringwald J, Rauh M,et al. Comparison of automated dif- ferential blood cell counts from Abbott Sapphire,Siemens Advia 120, Beckman Coulter DxH 800, and Sysmex XE-2100 in normal and pathologic samples[J]. Am J ClinPathol, 2013,139 ( 5 ) : 641-650.
  • 6Korninger L,Mustafa G,Schwarzinger I. The HaematologyAnalyser SF-3000 : Performance of the automated white blood cell differential count in comparison to the Haematology Analyser NE-1500[J]. Clin Lab Haematol, 1998,20(2) :81-86.
  • 7Vives-Corrons J L, Besson I ,Jou J M ,et al. Evaluation of the Abbott Cell-DYN 3500 hematology analyzer in university hospital[J]. Am J ClinPathol, 1996,105 (5) : 553-559.
  • 8Ruzicka K,Veitl M,Thalhammer-Scherrer R,et al. The new hema- tology analyzer SysmexXE-2100:Performance evaluation of a novel white blood cell differential technology[J]. Arch Pathol Lab Med, 2001,125(3) :391-396.
  • 9Kim S J, Kim Y, Shin S,et al. Comparison study of the rates of manu- al peripheral blood smear review from 3 automated hematology ana- lyzers, UnicelDxH 800, ADVIA 2120i, and XE 2100, using inter- national consensus group guidelines[J]. Arch Pathol Lab Med,2012, 136(11 ) : 1 408-1 413.

同被引文献13

引证文献3

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部