摘要
研究了长波8~15μm波段,阻值大于440QMCT光导红外探测器,探测率在10kHZ,14μm大于4×10如CiTI·Hz^1/2/W,在1kHZ和10kHz中心频率下的噪声测试,中波5~8μm红外光伏型InSb器件,探测率在25kHz,8.26μm大于l×10ucm·Hz1/2/W,在1kHZ和255kHZ中心频率下的噪声测试,并对器件信号进行了测试。信号和噪声测试是在124A锁相放大器测试系统测试,对124A测试系统的不确定度进行了分析,并与动态信号分析仪35670A对器件在0~50kHZ频谱范围的噪声进行了测试和比较。实验结果表明,高阻值的光导器件在1kHz和10kHz中心频率下噪声相差约1.4倍,光伏型InSb器件在1kHZ和15kHz中心频率下噪声相差约1.5倍,信号测试结果在1kHZ下和3kHz中心频率下变化不超过3%。通过测试和比较,对高频下的测试给出了建议。
The photo conductive detector in 8-15 lam band, resistance quality greater-than-equal match 440 Ω, detectivity(10 kHz, 14μm) greater-than-equal match 4× 10^10 cm . Hz^1/2/W, whose noise test in 1 kHz and 10 kHz frequency of center of lock-in 124A amplifier is studied in this paper. InSb photovoltaic detector in media wave of 5--8/,m, deteetivity(25 kHz, 8.26/xm) greater-than-equal match 1× 10Hcm .Hz^1/2/W, whose noise in 1 kHz and 25 kHz frequency of center of lock-in 124A amplifier is tested. The signal of detector and noise are also tested in 1 kHz,2 kHz,3 kHz center frequency of lock-in 124A amplifier, system test inaccuracy of 124A is analyized. The noise test result of 124A is compared with 35670A dynam ic signal analyzer test result. The test results approve that the noise in 1 kHz is 1.5 mutiple in 10 kHz of photo conductive detector, 1.5 multiple in 15 kHz of photovoltaic detector. The signal test result in 1 kHz and 3 kHz center frequency of 124A vary less than 3 percentage. Finally this paper gives advise to high frequency test through analysis of test dada.
出处
《光学与光电技术》
2014年第2期19-22,共4页
Optics & Optoelectronic Technology
关键词
红外探测器
噪声测试
高频测试
动态信号分析仪
infrared detector
noise test
high frequency test, dynamic signal analyzer