期刊文献+

医用50MHz超声换能器材料厚度影响仿真研究 被引量:8

Simulation Study on the Effect of Thickness for 50MHz Medical Ultrasound Transducer
下载PDF
导出
摘要 通过有限元分析法对高频超声换能器各组成部分的厚度影响进行了系统分析。创建了一个合理的有限元模型,通过仿真得到了不同条件下的声场特性与振动特性。结果表明,匹配层可改善声波透射的穿透特性并对导纳曲线变化有较大影响,引入了另一个共振频率;且厚40μm的压电振片的最优匹配层厚度是13.5μm。背衬层的厚度效应分析结果证明背衬层可改善声场的特性,一定程度上改善了导纳曲线,但背衬层的厚度变化对声场的影响较小。 In this paper, the effect of thickness of each component for 50 MHz ultrasound transducer was inves- tigated based on the finite element model (FEM). A reasonable finite element model has been established, and the sound fields and vibration characteristics in different conditions have been obtained by simulation. The results show that adding a match layer can improve the sound penetration transmission and has great influence on the admittance curve on which another resonant frequency appears; and the optimum thickness of match layer is 13.5μm for 40μm piezoelectric disc. The analysis of effect of thickness of backing layer shows that it can improve the sound field char- acteristic and the admittance curve at some extent, but the change of its thickness has small influence on the sound field
出处 《压电与声光》 CSCD 北大核心 2014年第2期205-209,216,共6页 Piezoelectrics & Acoustooptics
基金 江苏省基础研究基金资助项目(BK2011332) 苏州市国际科技合作计划基金资助项目(SH201102) 国家科技支撑计划基金资助项目(2012BAI13B02)
关键词 高频超声换能器 厚度影响 仿真 high frequency ultrasound transducer effect of thickness simulation
  • 相关文献

参考文献10

  • 1LETHIECQ M, LEVASSORT F,CERTON D,et al. Piezoelectric transducer design for medical diagnosis and NDE, piezoelectric and acoustic materials for transducer applications[M]. New York, USA,Springer Science+ Business Media, LLC, 2008.
  • 2MEITZLER A. TIERSTEN H F. WARNER A W. et al. IEEE standard on piezoelectricity [S]. New York: The Institute of Electrical and Electronics Engineers. 1988.
  • 3万明习,宗瑜瑾,王素品.生物医学超声学[M].北京:科学出版社,2010:652-654.
  • 4GUO N .CA WLEY P. Transient response of piezoelectric discs to applied voltage pulses [J]. Ultrasonic. 1991.29(1):208-217.
  • 5GUO N,C CAWLEY P. The finite element analysis of the vibration characteristics of piezoelectric discs[J]. Journal of Acoustic and Vibration .1992.159 (1): 115- 138.
  • 6ASSAAD J. MARTINE R. CHRISTIAN B.et al. Influence of the thickness and the attenuation coefficient of a backing on the response of transducers[J]. Ultrasonics .1996.34: 103-106.
  • 7ABBOUD N N. GREGORY L W. DA VID K V. et al. Finite element modeling for ultrasonic transducers [C]/ /San Diego. USA: Proc SPIE Int Syrnp Medical Imaging. 1998 .
  • 8LORENZO S, MARCO C. Design and optimization of a high performance ultrasound imaging probe through FEM and KLM Models[ CJ/ /Bo510n. USA: COMSOL Conference, 2011.
  • 9YUAN J R,JIANG X,CAO Peijie,et al. High frequency piezo composites micro fabricated ultrasound transducers for intravascular imaging[C]/ /VanccuvervCanada: IEEE Ultrasonics Symposium ,2006.
  • 10JONATHAN M C,TIMOTHY A R.CHEN W H,et al, Design of efficient. broadband single-element (20- 80 MHz) ultrasonic transducers for medical imaging applications [J]. IEEE Transactions on Ultrasonic, Ferroelectrics , and Frequency Control. 2003,50(1l): 1548-1557.

共引文献13

同被引文献58

  • 1张仲彬,李煜,杜祥云,董兵,李兴灿,徐志明.水质对板式换热器结垢的影响权重及其机制分析[J].中国电机工程学报,2012,32(32):69-74. 被引量:17
  • 2张烈辉,李允.低渗透气藏流——固耦合渗流数值模拟[J].天然气工业,2004,24(10):80-82. 被引量:7
  • 3彭贵勇,代建华,房殿春,李向红.内镜超声在消化道黏膜下肿瘤诊断与治疗中的价值[J].中华消化内镜杂志,2006,23(2):102-105. 被引量:92
  • 4钟云晴.根据不同模量弹性理论设计铸铁液压缸壁厚[J].机械设计与制造,2007(9):121-123. 被引量:1
  • 5Madsen EL, Frank GR, McCormick MM, et al. Anechoic sphere phantoms for estimating 3-D resolution of very-high-frequency ultr- asound scanners[J]. IEEE Trans Ultrason Ferr, 2010, 57(10): 2284- 2292.
  • 6Qiu WB, Yu YY, Tsang FK, et al. An FPGA-based open platform for ultrasound biomicroscopy[J]. IEEE Trans Ultrason Ferr, 2012, 59(7): 1432-1442.
  • 7胡袜,傅剑华,戎铁华,等.超声内镜检查在食管癌术前临床分期的应用价值[J].ChineseJoumalofCanceL200524(11):1358一1362.
  • 8Hasegawa N, Niwa Y, Arisawa T, et al. Preoperative staging of superficial esophageal carcinoma: comparison of an ultrasound probe and standard endoscopic ultrasonography.[J]. Gastrointestinal Endoscopy, 1996, 44(4):388-393.
  • 9Nishimaki T. Tanaka O, Ando N, et al. Evaluation of the accuracy of preoperative staging in thoracic esophageal cancer[J]. Annals of Thoracic Surgery, 1999, 68(6):2059-2064.
  • 10Massari M, Pavoni G M, Cioffi U, et al. [The rote of echo-endoscopy in the staging of squamous-cell carcinoma of the esophagus. The correlation between the surgical and anatomicopathological findings].[J]. Minerva Chirurgica, 1999, 54(4):205-12.

引证文献8