期刊文献+

主从式遥操作血管介入机器人 被引量:5

Master-Slave Remote Controlled Vascular Interventional Robot
下载PDF
导出
摘要 传统的血管介入手术需要医生徒手操控导管,存在医生受辐射和人为因素影响手术质量等弊端.针对该问题,设计了用于辅助医生手术的主从式遥操作血管介入机器人.根据临床手术的要求,基于模块化理念,分别对机器人的从端推进机构、主端操控装置进行了结构设计,为主从模式的实现做了相关的软、硬件开发;对机器人样机进行了必要的性能测试,证实了该机器人具有小于1 mm的运动控制精度和0.44 ms的主从运动实时性.结果表明:该机器人系统解决了传统手术中的问题,在功能和性能上完全满足血管介入手术的要求,具有较高的临床使用价值. Catheter should be controlled by doctor' s hands during the traditional vascular interventional surgery which had some disadvantages, as the doctor absorbed radiation and operation quality was affected by human factors. To solve the problem, master-slave remote controlled vascular interventional robot assisting doctor was designed. According to the requirements of clinical surgery, the subordinate feed mechanism and master operational device were designed respectively on the basis of the modular concept, the hardware and software were developed to achieve the master-slave mode. The necessary performances were tested, the motion control accuracy of less than 1 mm and the master-slave real-time motion of 0.44 ms were confirmed. The results showed that the robot system can solve the problems of traditional surgery, and meet the requirements of vascular interventional surgery on function and performance, which has high value for clinical application.
出处 《东北大学学报(自然科学版)》 EI CAS CSCD 北大核心 2014年第4期569-573,共5页 Journal of Northeastern University(Natural Science)
基金 国家高技术研究发展计划项目(2009AA044002)
关键词 血管介入 机器人 临床手术 控制精度 实时运动 vascular intervention robot clinical surgery control accuracy real-time motion
  • 相关文献

参考文献10

  • 1Ma X, Guo S X, Xiao N, et al. Development of a novel robot-assisted cather system with force feedback [ C]// Proceeding of the 2011 IEEE International Conference on Mechatronics and Automation. Beijing ,2011 : 107 - 111.
  • 2Wang T M, Zhang D P, Liu D. Remote-controlled vascular interventional surgery robot[ J]. The International Journal of Medical Robotics and Computer Assisted Surgery, 2010,2 :194 -201.
  • 3卢旺盛,刘达,田增民,张大鹏.血管介入手术机器人的关键技术分析[J].生物医学工程研究,2009,28(4):303-306. 被引量:15
  • 4Mack M J. Minimally invasive and robotic surgery [ J ]. Journal of the American Medical Association,2001,285 (5) : 568 - 572.
  • 5段星光,陈悦,于华涛.微创血管介入手术机器人控制系统与零位定位装置设计[J].机器人,2012,34(2):129-136. 被引量:8
  • 6Beyar R, Gruberg L, Roguin A, et al. Remote-control percutaneous coronary interventions [ J 1. Journal of American College of Cardiology, 2006,47 ( 2 ) :296 - 300.
  • 7Saliba W, Vivek Y. Atrial fibrillation ablation using a robotic catheter remote control system: initial human experience and long-term follow-up results [ J ]. Journal of the American College of Cardiology, 2008,51 ( 24 ) : 2407 - 241.
  • 8Guo S X, Kondo H, Wang J, et al. A new catheter operating system for medical applications [ C ]//International Conference on Complex Medical Engineering. Beijing: IEEE/ICME, 2007 : 82 - 86.
  • 9Saliba W, Cummings J E, Zhang Y, et al. A novel robotic catheter remote control system [ J 1. Feasibility and Safety of Transseptal Puncture and Endocardial Catheter Navigation, 2006,17(10) :1102 - 1105.
  • 10王益红,陈志同.基于PMAC的数控机床手轮功能的实现[J].机械工程师,2005(12):68-70. 被引量:8

二级参考文献11

共引文献28

同被引文献28

引证文献5

二级引证文献13

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部