去金属伪影序列在脊柱金属植入物患者MR成像中的应用 被引量:27

Clinical application of syngo WARP in the metallic prostheses MR imaging
原文传递
导出
摘要 目的 探讨去金属伪影(syngo WARP)序列在金属植入物患者脊柱MRI中的应用价值.方法 收集25例脊柱(颈、腰椎)金属植入物患者,均行脊柱MR常规序列及syngo WARP优化序列扫描,分析2种序列检查的扫描时间及图像质量评分,并采用配对t检验进行比较.结果 所有患者均顺利完成检查.颈椎常规序列检查时间为8 min 16 s,syngo WARP优化序列检查时间为12 min 45 s,差异有统计学意义(t=7.963,P<0.01);腰椎常规序列和syngo WARP优化序列的检查时间分别为9 min 16 s和11 min 53 s,差异有统计学意义(t=4.904,P<0.01).syngo WARP优化序列图像质量评分为(4.22±0.67)分,高于常规序列的(3.56±0.53)分,差异有统计学意义(t=3.364,P=0.002).syngo WARP优化序列图像上金属植入物周边软组织信号丢失减轻,图像变形与模糊情况都较常规序列明显改善.结论 syngo WARP技术可以减轻金属植入物MR伪影,改善图像质量,在金属植入物患者脊柱MRI中具有较大的应用价值. Objective To investigate the value of syngo WARP technology for spine MR imaging in patients with metallic prosthesis.Methods Twenty-five patients with metallic prosthesis in cervical spine or lumbar spine were prospectively included in this study.Both conventional MR sequences and optimized sequences with syngo WARP were applied in all patients.Acquisition time and image quality of two sequences were compared using a paired t test.Results All patients were examined successfully.Scanning time of cervical spine was 8 min 16 s vs.12 min 45 s for conventional sequences and syngo WARP optimized sequences respectively,with a significant difference (t =7.963,P < 0.01).Scanning time of lumbar spine was longer by syngo WARP optimized sequences compared with conventional sequences (11 min 53 s vs.9 min 16 s),with a significant difference (t =4.904,P < 0.01).However,image quality was better for syngo WARP optimized sequences compared with conventional sequences (4.22 ± 0.67 vs.3.56 ± 0.53 ;t =3.364,P =0.002).STIR with syngo WARP could optimize the signal loss of metallic prosthesis surrounding tissues,and significantly improve image distortion and blurring compared to the conventional sequence.Conclusion Syngo WARP technique could effectively reduce metal artifacts and better display metal implant surrounding tissue and anatomical structure with potential clinical value.
出处 《中华放射学杂志》 CAS CSCD 北大核心 2014年第4期320-323,共4页 Chinese Journal of Radiology
关键词 金属 图像质量 磁共振成像 Metals Image quality Magnetic resonance imaging
  • 相关文献

参考文献17

  • 1曹筝,陈莉丽,龚向阳.口腔金属材料磁共振成像伪影的初步探讨[J].中华医学杂志,2008,88(26):1855-1858. 被引量:7
  • 2Sutter R, Ulbrich E J, Jellus V, et al. Reduction of metal artifacts in patients with total hip arthroplasty with slice-encoding metal artifact correction and view-angle tilting MR imaging [J]. Radiology, 2012, 265:204-214.
  • 3Darnell JC, Saltiel AR. Coenzyme A-dependent, ATP- independent acylation of 2-acy! lysophosphatidylinositol in rat liver microsomes[ J]. Biochim Biophys Acta, 1991, 1084:292-299.
  • 4Knight SR, Aujla R, Biswas SP. Total Hip Arthroplasty-over 100 years of operative history[ J]. Orthop Rev ( Pavia), 2011,3 : e16.
  • 5Learmonth ID, Young C, Rorabeck C. The operation of the century: total hip replacement [ J ]. Lancet, 2007, 370: 1508-1519.
  • 6Goodman S. Wear particulate and osteolysis [ J ] . Orthop Clin North Am ,2005,36:41-48.
  • 7Gallo J, Goodman SB, Konttinen YT, et al. Osteolysis around total knee arthroplasty : a review of pathogenetic mechanisms [ J ].Acta Biomater, 2013,9 : 8046-8058.
  • 8Anderson H, Toms AP, Cabir JG, et al. Grading the severity of soft tissue changes associated with metal-on-metal hip replacements: reliability of an MR grading system [ J ]. Skeletal Radiol, 2011,40:303-307.
  • 9Campbell P, Ebramzadeh E, Nelson S, et al. Histological features of pseudotumor-like tissues from metal-on-metal hips [ J ]. Clin Orthop Relat Res, 2010,468 : 2321-2327.
  • 10Counsell A, Heasley R, Arumilli B, et al. A groin mass caused by metal particle debris after hip resurfacing [ J ]. Aeta Orthop Belg, 2008,74:870-874.

二级参考文献11

  • 1吴胜勇,尉传社.3.0TMR成像系统的优势及其在中枢神经系统中的应用[J].医学综述,2005,11(1):93-95. 被引量:4
  • 2周诚.重视CT、MRI腹部检查技术的应用[J].中华医学杂志,2005,85(5):289-290. 被引量:6
  • 3Gray CE, Redpath TW, Smith FW, et al. Advanced imaging: magnetic resonance imaging in implant dentistry: a review. Clin Oral Implants Res, 2003, 14: 18-27.
  • 4Kuhl CK, Traber F, Schild HH, et al. Whole-body high-field-strength (3. 0-T ) MR imaging in clinical practice part Ⅰ. Technical considerations and clinical applications. Radiology, 2008, 246:675-696.
  • 5Lissac M, Metrop D, Brugirard J , et al. Dental materials and magnetic resonance imaging. Invest Radiol, 1991,26:40-45.
  • 6Hubalkova H, Hora K, Seidl Z, et al. Dental materials and magnetic resonance imaging. Eur J Prosthodont Restor Dent, 2002, 3: 125-130.
  • 7Abbaszadeh K, Heffez LB, Mafee MF, et al. Effect of interference of metallic objects on interpretation of T1 -Wighted magnetic resonance images in the maxillofacial region. Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 2000, 89:759-765.
  • 8Burtscher IM, Owman T, Romner B, et al. Aneurysm clip MR artifacts: titanium versus stainless steel and influence of imaging parameters. Acta Radiol, 1998, 39 : 70-76.
  • 9Olsrud J, Latt J, Brockstedt S, et al. Magnetic resonance imaging artifacts caused by aneurysm clips and shunt valves: dependence on field strength ( 1.5 and 3 T) and imaging parameters. J Magn Reson Imaging, 2005, 22:433-437.
  • 10Meindl T, Born C, Britsch S, et al. Functional BOLD MR.I: comparison of different field strengths in a motor task. Eur Radiol, 2008, 18:1102-1113.

共引文献6

同被引文献149

  • 1戈明媚,王秋良,刘志钦,刘晓林,廖玉珍,孙玮,孙楠,张英魁.MR螺旋桨扫描技术在消除伪影方面的临床应用[J].中华放射学杂志,2006,40(2):208-212. 被引量:37
  • 2黄群,杨海涛,李维金,夏黎明,王仁法,王承缘.脊柱内固定术后MRI应用研究[J].放射学实践,2007,22(7):743-746. 被引量:5
  • 3霍沃德里.颅脑MR和CT诊断学[M].西安:世界图书出版西安公司,2001:75-99.
  • 4康庄,卢广文,邹艳,王亮.骨关节不锈钢钉植入术后的磁共振影像技术[J].南方医科大学学报,2007,27(8):1282-1284. 被引量:2
  • 5曹玉净,沈云辉,艾进伟.手法复位结合中药熏洗对踝关节骨折患者等速肌力的影响[J].上海中医药杂志,2007,41(9):65-66. 被引量:19
  • 6Sampson, B.,Hart, A..Clinical usefulness of blood metal measurements to assess the failure of metal-on- metal hip implants[J].Annals of Clinical Biochemistry: Journal of the Association of Clinical Biochemists in Association with de Nederlandse Vereniging voor Klinische Chemie, 2012,49(2):118-131.
  • 7Stefano Mischler, Anna Igual Munoz. Wear of CoCrMo alloys used in metal-on-metal hip joints: h tribocorrosion appraisal [J]. Wear: an International Journai on the Science and Technology of Friction, Lubrication and Wear, 2013,297(12):1081-1094.
  • 8高元贵,蔡幼拴,蔡祖龙.磁共振成像诊断学[M].3版.北京:人民军医出版社.1999:122-128.
  • 9Shellock FG, Shellock VJ. Metallic stents: evaluation of MR imaging safety[J]. AJR Am J Roentgenol, 1999, 173 (3):543-547.
  • 10Viano AM, Gronemeyer SA, Haliloglu M, et al. Improved MR imaging for patients with metallic implants[J]. Magn Reson Imaging, 2000, 18(3):287-295.

引证文献27

二级引证文献90

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部