vHP法在主动脉、冠状动脉及肺动脉一站式成像中的应用研究 被引量:3

Application of variable helical pitch for whole imaging of aorta,coronary and pulmonary arteries
下载PDF
导出
摘要 目的:探讨320排容积CT可变螺距法(vHP)在急性胸痛患者全主动脉、冠状动脉及肺动脉一站式成像中的临床应用价值。方法:以30例不明原因的急性胸痛患者为实验组,行320排容积CT vHP法扫描:主动脉弓上缘至心底水平采用小螺距心电门控,心底至髂动脉水平采用大螺距不加心电门控。以30例非心电门控大螺距全主动脉320排容积CT扫描患者作为对照组,比较两组的图像质量、辐射剂量以及对比剂使用量。结果:实验组图像对主动脉瓣及主动脉根部的可诊断率(100%,30/30)明显高于对照组(30%,9/30);所有对照组图像对冠状动脉均无法评价,实验组冠状动脉图像的可诊断率达98%;两组图像对肺动脉的可诊断率差异无统计学意义(P>0.05)。实验组每例患者的辐射剂量[(1131.73±58.08)mGy·cm]稍高于对照组[(1102.67±70.29)mGy·cm],两组间差异无统计学意义(P>0.05);实验组中每例患者的对比剂用量[(96.07±5.62)mL]略高于对照组[(76.00±5.96)mL],差异有统计学意义(P<0.05)。结论:vHP扫描法可以在不明显增加辐射剂量的前提下一次完成急性胸痛患者主动脉、冠状动脉及肺动脉的扫描,并可减少对比剂的用量,对于进行胸痛患者尽早明确诊断有一定的应用价值。 Objective:To explore the value of whole imaging of aorta,coronary arteries,and pulmonary arteries by u- sing variable helical pitch (vHP) method of 320-slice volume CT. Methods: 30 unexplained patients with acute chest pain from our hospital were grouped as the experimental group by vHP scanningl Small pitch ECG-gating CT angiography were performed from the upper edge of the aortic arch to the bottom of the heart, while big pitch without ECG-gating were per formed from the bottom of the heart to the level of iliac artery. The other 30 patients who were scanned by big pitch without ECG-gating for the whole aorta were placed as the control group. The image quality, radiation dose and contrast medium were recorded, and the statistical analysis was performed. Results:The diagnosis rate of the aortic valve and root for the experimental group (100% ,30/30) was higher than that of the control group (30%,9/30). All the coronary arteries of the control group could not be evaluated,and the diagnosis rate of coronary arteries for the experimental group was 98%. There was no significant difference for the diagnosis of pulmonary arteries and radiation dose between the two groups. The contrast medium volume of the experimental group [(96.07±5.62)mL] was higher than that E(76.00±5.96)mL] of the control group (P〈0.05). Conclusion:The vHP method can obtain the whole image of aorta, coronary, and pulmonary arte ries by only one scan without obvious increasing of the radiation dose, and the volume of contrast medium can be reduced, so it can be used as a valuable method for the early diagnosis of acute chest pain.
出处 《放射学实践》 2014年第4期400-403,共4页 Radiologic Practice
关键词 可变螺距 急性胸痛 体层摄影术 X线计算机 血管成像 一站式成像 冠状动脉 肺动脉 Variable helical pitch Acute chest pain Tomography,X-ray computed Angiography One stop ima-ging Coronary artery Pulmonary artery
  • 相关文献

参考文献13

  • 1Rubin GD,Shiau MG,Leung AN,et al.Aorta and iliac arteries:single versus multiple detector-row helical CT angiography [J].Radiology,2000,215(3):670-676.
  • 2Bax JJ,Schuijf JD.Which patients should be referred for non-inva-sive angiography with multi-slice CT[J].Int J Cardiol,2007,114(1):1-3.
  • 3Fujioka C,Horiguchi J,Kiguchi M,et al.Survey of aorta and coro-nary arteries with prospective ECG-triggered IOOkV 64-MDCTangiography[J].AJR,2009,193(1):227-233.
  • 4Rybicki FJ,Otero HJ,Steigner ML,et al.Initial evaluation ofcoronary images from 320-detector row computed tomography[J].Int J Cardiovasc Imaging,2008,24(5):535-546.
  • 5Yoon YE,Wann S.Evaluation of acute chest pain in the emergen-cy department?" triple rule-out" computed tomography angiogra-phy[J].Cardiol Rev,2011,19(3):115-121.
  • 6Ko SF,Hsieh MJ,Chen MC,et al.EffectvS of heart rate on motionartifacts of the aorta on non-ECG-assisted 0.5s thoracic MDCT[J].AJR,2005,184(4):1225-1230.
  • 7Boiselle PM.Computed tomography screening for lung cancer[J].JAMA,2012,309(11):1163-1170.
  • 8Brenner DJ,Hall EJ.Computed tomography-an increasingsource of radiation exposure[J].N Engl J Med,2007,357(22):2277-2284.
  • 9刘一,晏子旭,吴晓峰,张兆琪.螺距的选择对胸腹主动脉夹层320层CT成像质量的影响[J].放射学实践,2011,26(3):350-353. 被引量:16
  • 10孙凯,韩瑞娟,马利军,王利军,王志清,郭城,赵冬梅,王刚.大螺距双源CT前瞻性心电门控扫描评价冠状动脉狭窄的临床研究[J].中华放射学杂志,2012,46(1):49-53. 被引量:44

二级参考文献12

  • 1Alkadhi H, Stolzmann P, Desbiolles L, et al. Low-dose, 128- slice, dual-source CT coronary angiography: accuracy and radiation dose of the high-pitch and the step-and-shoot mode. Heart, 2010,96:933-938.
  • 2Achenbach S, Marwan M, Ropers D, et al. Coronary computed tomography angiography with a consistent dose below 1 mSv using prospectively electrocardiogram-triggered high-pitch spiral acquisition. Eur Heart J, 2010,31:340-346.
  • 3Austen WG, Edwards JE, Frye RL, et al. A reporting system on patients evaluated for coronary artery disease, report of the ad hoc committee for grading of coronary artery disease, council on cardiovascular surgery, american heart association. Circulation, 1975,51 (4 Suppl) :5-40.
  • 4Karlo C, Leschka S, Goetti RP, et al. High-pitch dual-source CT angiography of the aortic valve-aortic root complex without ECG- synchronization. Eur Radiol, 2011,21:205-212.
  • 5Ropers U, Ropers D, Pflederer T, et al. Influence of heart rate on the diagnostic accuracy of dual-source computed tomography coronary angiography. J Am Coll Cardiol, 2007,50: 2393-3938.
  • 6Leschka S, Stolzmann P, Desbiolles L, et al. Diagnostic accuracy of high-pitch dual-source CT for the assessment of coronary stenoses : first experience. Eur Radiol, 2009,19:2896-2903.
  • 7Raft GL, Gallagher MJ, O'Neill WW, et al. Diagnostic accuracy of noninvasive coronary angiography using 64-slice spiral computed tomography. J Am Coil Cardiol, 2005,46:552-557.
  • 8Scheffel H, Alkadhi H, Plass A, et al. Accuracy of dual-source CT coronary angiography: first experience in a high pre-test probability population without heart rate control. Eur Radiol, 2006,16:2739-2747.
  • 9Scheffel H, Alkadhi H, Leschka S, et al. Low-dose CT coronary angiography in the step-and- shoot mode: diagnostic performance. Heart, 2008,94 : 1132-1137.
  • 10尚建敏,夏进东,徐寅学,陈晓蓉,曹院国,万华,罗春霞,宋亚宾.16层螺旋CT血管成像技术在主动脉夹层中的临床价值[J].放射学实践,2008,23(11):1212-1215. 被引量:9

共引文献58

同被引文献21

引证文献3

二级引证文献28

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部