摘要 目的探讨K锉初步预备和扭矩对机用ProTaper抗折性能的影响。方法分别计算各组机用ProTaper器械折断或变形前所预备的S型树脂模拟根管的数目。结果 A1组(模拟根管未经K锉初步预备)S1预备了13个模拟根管;B1组(模拟根管经K锉初步预备)S1预备了36个模拟根管;A2组(低扭矩组)S2预备了17个模拟根管;B2组(高扭矩组)S2预备了18个模拟根管;A3组(低扭矩组)F1预备了14个模拟根管;B3组(高扭矩组)F1预备了14个模拟根管;A4组(低扭矩组)F2预备了6个模拟根管;B4组(高扭矩组)F2预备了6个模拟根管。结论根管经K锉初步预备可降低机用ProTaper器械折断的发生;扭矩的高低对机用ProTaper器械的抗折性能没有明显影响。 Objective This study aims to investigate the influence of K file preflaring and torque on the failure of rotary nickel-titanium ProTaper instruments. Methods An in vitro method was carried out to calculate the mean number of S-shaped simulated root canals before file fracture or deformation under different conditions. Results Group AI(S1 on simulators with no preflaring) shaped 13 blocks, and Group Bl(S1 on manually preflared simulators) shaped 36 blocks. Group A2(S2 with low torque) shaped 17 blocks, and Group B2(S2 with high torque) shaped 18 blocks. Group A3 (F1 with low torque) shaped 14 blocks, and Group B3(F1 with high torque) shaped 14 blocks. Group A4 (F2 with low torque) shaped 6 blocks, and Group B4(F2 with high torque) shaped 6 blocks. Conclusion K file preflaring is a major determinant in reducing the failure of rotary nickel-titanium ProTaper instruments, whereas torque exhibiting no influence on the failure of rotary nickel-titanium ProTaper instruments.
出处 《国际口腔医学杂志》 CAS 2014年第3期289-291,共3页 International Journal of Stomatology
关键词 K锉 扭矩 ProTaper器械 K file torque ProTaper rotary instruments
  • 相关文献

参考文献5

  • 1Wu J,Lei G,Yan M,et al.Instrument separation analysis of multi-used ProTaper Universal rotary system during root canal therapy[J].J Endod,2011,37(6):758-763.
  • 2Gutmann JL,Gao Y.Alteration in the inherent metallic and surface properties of nickel-titanium root canal instruments to enhance performance,durability and safety:a focused review[J].Int Endod J,2012,45(2):113-128.
  • 3彭彬,沈雅,樊明文.牙体牙髓临床治疗 Ⅳ.镍钛合金根管器械折断的临床特点及防治[J].中华口腔医学杂志,2006,41(5):309-312. 被引量:27
  • 4Yared GM,Bou Dagher FE,Machtou P.Influence of rotational speed,torque and operator's proficiency on ProFile failures[J].Int Endod J,2001,34(1):47-53.
  • 5Berutti E,Negro AR,Lendini M,et al.Influence of manual preflaring and torque on the failure rate of ProTaper rotary instruments[J].J Endod,2004,30(4):228-230.

二级参考文献10

  • 1Sattapan B,Nervo GJ,Palamara JE,et al.Defects in rotary nickel titanium files after clinical use.J Endod,2000,26:161-165.
  • 2Cheung GS,Peng B,Bian Z,et al.Defects in ProTaper S1 instruments after clinical use:fractographic examination.Int Endod J,2005,38:802-807.
  • 3Pruett JP,Clement DJ,Carnes DL Jr.Cyclic fatigue testing of nickel-titanium endodontic instruments.J Endod,1997,23:77-85.
  • 4Peng B,Shen Y,Cheung GS,et al.Defects in ProTaper S1 instruments after clinical use:longitudinal examination.Int Endod J,2005,38:550-557.
  • 5Shen Y,Peng B,Cheung GS.Factors associated with the removal the fractured NiTi instruments from root canal systems.Oral Surg Oral Med Oral Pathol Oral Radiol Endod,2004,98:605-610.
  • 6Shen Y,Cheung GS,Bian Z,et al.Comparison of defects in ProFile and ProTaper systems after clinical use.J Endod,2006,39:61-65.
  • 7Ward JR,Parashoes P,Messer HH.Evaluation of an ultrasonic technique to remove fractured rotary nickel-titanium endodontic instruments from root canals:an experimental study.J Endod,2003,29:756-763.
  • 8Pitt Ford TR,Rhodes JS,Pitt Ford HE.Endodontics:problemsolving in clinical practice.London:Martin Dunitz Ltd,2002.137-148.
  • 9Ruddle CJ.Non-surgical endodontic retreatment.In:Cohen S,Burns RC.Pathways of the pulp.8th ed.St Louis:Mosby,2002.875 -929.
  • 10沈雅,彭彬,范兵,边专,樊明文.镍钛合金根管器械折断的临床分析[J].中华口腔医学杂志,2004,39(1):38-41. 被引量:51