摘要
EILog-05岩性密度仪器测井时存在测量误差较大、重复性、稳定性较差等问题。根据误差影响因素分析的结论,对仪器源距部分进行了重新设计。通过仪器刻度校验数据、测井试验结果对比分析,仪器测量性能有了很大改善,实现了优化设计的目的。
EILog-05 litho-density logging tool has the problems of larger measurement error,poor repeatability and stability.According the error influencing factor analysis conclusion,we redesign the source detector separation part of the logging tool.The comparative analysis of instrument calibration data and logging test result show the tool performance improves a lot,realizing the aim of optimal design.
出处
《石油仪器》
2014年第1期51-53,100,共3页
Petroleum Instruments