工业CT探测尺寸误差的校准及误差分析 被引量:12

Calibration and Analysis for Probing Size Error of Industrial CT
下载PDF
导出
摘要 针对工业CT探测尺寸误差的校准需求,提出一套基于标准球的校准工业CT探测尺寸误差的方法,并对引起探测尺寸误差的各种因素进行了详细分析。首先使用中国计量科学研究院标准测长机对标准球的直径进行校准,再使用标定后的标准球校准工业CT探测尺寸误差。实验结果表明球标准器能够满足工业CT探测尺寸误差校准的需求,既能被精密校准,又可在工业CT中获得低伪像的高质量图像,从而实现对工业CT探测尺寸误差的校准。通过对影响工业CT探测尺寸误差因素的分析,讨论了工业CT成像过程中各环节的误差源,对于建立工业CT测量系统全面的校准方法和溯源具有重要意义。 According to the needs of calibrating the probing size error, a method based on the standard balls for the calibration of industrial CT probing size error is proposed and factors which cause the probing size error are detailly analyzed. Firstly, it calibrates the diameter by the length-measuring machine of National Institute of Metrology, then calibrates the industrial CT probing size error by these balls. The experimental results show that these standard balls can be calibrated accurately, and high quality image with lower artifacts can be acquired. And this kind of standards can meet the calibration needs of probing size error of industrial CT system. Error sources in imaging process, which influence the probing size error are discussed and analyzed. It is of great significance to establish calibration and traceability method for industrial CT system.
出处 《计量学报》 CSCD 北大核心 2014年第3期216-220,共5页 Acta Metrologica Sinica
基金 基金项目:国家重大仪器专项基金(2011YQ03011208)
关键词 计量学 工业CT 标准球 探测尺寸误差 Metrology Industrial CT Standard ball Probing size error
  • 相关文献

参考文献11

  • 1Bartscher M, Neuschaefer-Rube U, Wiildele F. Computed tomography-a highly potential tool for industrial quality control and production near measurements [ J ] . CIRP Annals-Manufacturing Technology, 2004, (1860) :477 -796.
  • 2Bartscher M, Neuschaefer-Rube U, Staude A, et al. Application of an industrial CT reference standard for cast free-form shaped work pieces [ C ]// International Symposium on Digital Industrial Radiology and Computed Tomography. Berlin, Germany, 2011, 20-25.
  • 3Voet A, Kiekens K, Welkenhuyzen F, et al. A test object for calibration and accuracy assessment in X-ray CT metrology [ C ] // 10th International Symposium on Measurement and Quality Control 2010. Osaka, Japan, 2010.
  • 4陈思,陈浩,李敬.一种工业CT测量精度评估方法[C]//全国射线数字成像与CT新技术.四川绵阳,2012,189-197.
  • 5Brunke O, Suppes A. Precise 3D dimensional metrology using high resolution X-ray computed tomography (TCT) [C]// SPIE. San Diego, California, 2010, (7804): 780400 - 1.
  • 6Prufstelle Testing and Certification Institute. VDI/VDE 2630-1. 3 [ S ]. Computed Tomography in dimensional measurement. 2011.
  • 7朱训生,任新生,杨建国,薛秉源.球度最小二乘评定的一般公式[J].上海交通大学学报,1998,32(5):43-45. 被引量:6
  • 8杨倩,潘晋孝.工业CT中机械定位误差对投影数据的影响[J].测试技术学报,2009,23(增刊):138-143.
  • 9王珏,陈教泽,谭辉,王鑫.工业CT探测系统噪声特性研究[J].核电子学与探测技术,2010,30(7):929-934. 被引量:6
  • 10张峰,闫镔,李建新,李磊,包尚联.工业X-CT散射校正技术综述[J].CT理论与应用研究(中英文),2009,18(4):34-43. 被引量:7

二级参考文献40

共引文献19

同被引文献94

  • 1程银宝,陈晓怀,王中宇,王汉斌,李红莉,李亚茹.CMM形状测量任务的不确定度分析与评定[J].计量学报,2020,41(2):134-138. 被引量:17
  • 2邓小武,黄劭敏,祁振宇.CT模拟机的质量控制和质量保证检验[J].中国肿瘤,2004,13(9):546-550. 被引量:33
  • 3先武,李时光,王珏.最佳无损检测手段──工业CT技术的发展[J].光电工程,1995,22(4):51-58. 被引量:36
  • 4先武,李时光.工业CT技术[J].无损检测,1996,18(2):57-60. 被引量:10
  • 5孙灵霞,叶云长.工业CT技术特点及应用实例[J].核电子学与探测技术,2006,26(4):486-488. 被引量:39
  • 6国家质量技术监督局计量司.测量不确定度评定与表示指南[M].北京:中国计量出版社,2005.
  • 7陈思,陈浩,李敬.一种工业CT测量精度评估方法[C]//全国射线数字成像与CT新技术.四川绵阳,2012,189-197.
  • 8Schmitt R, Niggemann C. Uncertainty in measurement for x-ray-computed tomography using calibrated work pieces [J]. Measurement Science and Technology, 2010, 21 (5) : 054008.
  • 9Kruth J P, Bartscher M, Carmignato S, et al. Computed tomography for dimensional metrology [ J ]. CIRP Annals- Manufacturing Technology, 2011, 60(2) : 821 - 842.
  • 10Bartscher M, Neuschaefer-Rube U, Staude A, et al. Application of an industrial CT reference standard for cast free-form shaped work pieces [ C ]// International Symposium on Digital Industrial Radiology and Computed Tomography, Berlin, Germany, 2011, 20 - 22.

引证文献12