摘要
目的:在医用高频电刀的质量控制工作中掌握高频电刀的检测方法。方法:采用德国S.P.L公司的400型高频电刀检测仪对某院的高频电刀进行检测。结果:通过检测,得出高频电刀总合格率以及引起输出功率和高频漏电流不合格的原因。结论:必须定期对高频电刀进行检测,以提高其在临床使用中的安全性。
Objective To make clear the quality detection method of medical high-frequency electrotomc. Methods Type 400 detection instrument from Germany S.P.L company was used to test the high-frequency electrotome in some hospital.Results The total qualified rate of the high-frequency electrotome was found out, and the factors for unqualified output power and leak current were also determined. Conclusion Periodical detection of high-frequency electrotome has to be performed for its clinical safety.
出处
《医疗卫生装备》
CAS
2014年第4期82-83,105,共3页
Chinese Medical Equipment Journal
关键词
高频电刀
检测
合格率
high-frequency electrotome
detection
qualified rate