期刊文献+

医用高频电刀的质量检测技术与实践 被引量:11

Quality Detection Technology and Practice of Medical High-frequency Electrotome
下载PDF
导出
摘要 目的:在医用高频电刀的质量控制工作中掌握高频电刀的检测方法。方法:采用德国S.P.L公司的400型高频电刀检测仪对某院的高频电刀进行检测。结果:通过检测,得出高频电刀总合格率以及引起输出功率和高频漏电流不合格的原因。结论:必须定期对高频电刀进行检测,以提高其在临床使用中的安全性。 Objective To make clear the quality detection method of medical high-frequency electrotomc. Methods Type 400 detection instrument from Germany S.P.L company was used to test the high-frequency electrotome in some hospital.Results The total qualified rate of the high-frequency electrotome was found out, and the factors for unqualified output power and leak current were also determined. Conclusion Periodical detection of high-frequency electrotome has to be performed for its clinical safety.
机构地区 解放军
出处 《医疗卫生装备》 CAS 2014年第4期82-83,105,共3页 Chinese Medical Equipment Journal
关键词 高频电刀 检测 合格率 high-frequency electrotome detection qualified rate
  • 相关文献

参考文献8

二级参考文献21

  • 1刘新颖.高频电刀临床使用的安全措施[J].医疗卫生装备,2005,26(6):44-44. 被引量:17
  • 2段乔峰,高山,张学浩.关于高频手术设备的高频漏电流的探讨[J].中国医疗器械信息,2006,12(8):47-49. 被引量:6
  • 3杨秀琴 李珂 等.使用高频电刀应防手术野烧伤[J].中华护理杂志,1996,31(10):570-570.
  • 4总后勤部卫生部.急救装备[M].北京:解放军出版社,2005:121-140.
  • 5GB9706.4-92高频手术设备专用安全要求[S].
  • 6GB9706.4-1999医用电气设备第二部分:高频手术设备安全专用要求[S].
  • 7GB9706.1—2007医用电气设备第一部分:安全通用要求[S].北京:中国标准出版社,2007.
  • 8GB9706.4—2009医用电气设备第二部分:高频手术设备安全专用要求[S].北京:中国标准出版社,2009.
  • 9国家技术监督局.JJF1217—2009高频电刀校准规范[S].北京:中国标准出版社,2007.
  • 10潘朝伟.手术室高频电刀的安全使用[J].河南外科学杂志,2007,13(4):60-60. 被引量:10