期刊文献+

A Novel High Nitrogen Nickel-free Coronary Stents System: Evaluation in a Porcine Model 被引量:2

A Novel High Nitrogen Nickel-free Coronary Stents System: Evaluation in a Porcine Model
下载PDF
导出
摘要 Objective To study the safety of the novel high nitrogen nickel-free austenitic stainless steel bare metal stents (BMS) in a recognized porcine coronary model and to select a better grid structure of it. Methods Three types of stents were randomly implanted in different coronary arteries of the same pig: 316L stainless steel BMS (316L-BMS) (n=12), novel high nitrogen nickel-free stents Grid A (NF-A-BMS) (n=12) and novel high nitrogen nickel-free stents Grid B (NF-B-BMS) (n=12). In total, eighteen animals underwent successful random placement of 36 oversized stents in the coronary arteries. Coronary angiography was performed after 36 d of stents implantation. Nine animals were respectively sacrificed after 14 d and 36 d for histomorphologic analysis. 〈br〉 Results Quantitative coronary angiography (QCA) showed similar luminal loss (LL) in the three groups:(0.21±0.17) mm for 316L-BMS, (0.16±0.12) mm for NF-A-BMS, (0.24±0.15) mm for NF-B-BMS (P=0.05). Histomorphomeric analysis after 15 d and 36 d revealed that there was also no significant difference among the three groups in neointimal area (NA) with similar injury scores respectively. High magnification histomorphologic examination showed similar inflammation scores in the three groups, but NF-A-BMS group had poorer endothelialization scores compared with NF-B-BMS group, 2.00±0.63 vs. 2.83±0.41 (P=0.015) at 15 d, which also could be proved by the scanning electron microscope. However, the difference could not been observed at 36 d. Conclusion The novel NF-BMS showed similar safety as 316L-BMS during the short-term study. NF-B-BMS had better endothelialization than NF-A-BMS and this may owe to the specific strut units. Objective To study the safety of the novel high nitrogen nickel-free austenitic stainless steel bare metal stents (BMS) in a recognized porcine coronary model and to select a better grid structure of it. Methods Three types of stents were randomly implanted in different coronary arteries of the same pig: 316L stainless steel BMS (316L-BMS) (n=12), novel high nitrogen nickel-free stents Grid A (NF-A-BMS) (n=12) and novel high nitrogen nickel-free stents Grid B (NF-B-BMS) (n=12). In total, eighteen animals underwent successful random placement of 36 oversized stents in the coronary arteries. Coronary angiography was performed after 36 d of stents implantation. Nine animals were respectively sacrificed after 14 d and 36 d for histomorphologic analysis. 〈br〉 Results Quantitative coronary angiography (QCA) showed similar luminal loss (LL) in the three groups:(0.21±0.17) mm for 316L-BMS, (0.16±0.12) mm for NF-A-BMS, (0.24±0.15) mm for NF-B-BMS (P=0.05). Histomorphomeric analysis after 15 d and 36 d revealed that there was also no significant difference among the three groups in neointimal area (NA) with similar injury scores respectively. High magnification histomorphologic examination showed similar inflammation scores in the three groups, but NF-A-BMS group had poorer endothelialization scores compared with NF-B-BMS group, 2.00±0.63 vs. 2.83±0.41 (P=0.015) at 15 d, which also could be proved by the scanning electron microscope. However, the difference could not been observed at 36 d. Conclusion The novel NF-BMS showed similar safety as 316L-BMS during the short-term study. NF-B-BMS had better endothelialization than NF-A-BMS and this may owe to the specific strut units.
出处 《Biomedical and Environmental Sciences》 SCIE CAS CSCD 2014年第4期289-294,共6页 生物医学与环境科学(英文版)
关键词 Austenitic steel Bare metal stent ENDOTHELIALIZATION RESTENOSIS Stent thrombosis Austenitic steel Bare metal stent Endothelialization Restenosis Stent thrombosis
  • 相关文献

参考文献16

  • 1Roiron C, Sanchez P, Bouzamondo A, et al. Drug-eluting stents: an updated meta-analysis of randomised controlled trials. Heart, 2006; 92, 641-9.
  • 2Fattori R and Piva T. Drug-eluting stents in vascular intervention. Lancet, 2003; 361, 247-9.
  • 3Coogan TP, Latta DM, Snow ET, et al. Toxicity and carcinogenicity of nickel compounds. Crit Rev Toxicol, 1989; 19 341-84.
  • 4Uter W, Pfahlberg A, Gefeller O, et al. Risk factors for contact allergy to nickel - results of a multifactorial analysis. Contact Dermatitis, 2003; 48, 33-8.
  • 5Silber S, Windecker S, Vranckx P, et al. Unrestricted randomised use of two new generation drug-elutingcoronary stents: 2-year patient-related versus stent-related outcomes from the RESOLUTE All Comers trial. Lancet, 2011; 377, 1241-7.
  • 6Claessen BE, Caixeta A, Henriques JP, et al. Current status of the Xience V everolimus-eluting coronary stent system. Expert Rev Cardiovasc Ther, 2010; 8, 1363-74.
  • 7Kereiakes DJ, Cannon LA, Eeldman RL, et al. Clinical and angiographic outcomes after treatment of de novo coronary stenoses with a novel platinum chromium thin-strut stent: primary results of the PERSEUS (Prospective Evaluation in a Randomized Trial of the Safety and Efficacy of the Use of the TAXUS Element Paclitaxel-Eluting Coronary Stent System) trial. J Am Coll Cardiol, 2010; 56, 264-71.
  • 8Schwartz RS, Huber KC, Murphy JG, et al. Restenosis and the proportional neointimal response to coronary artery injury: Results in a porcine model. J Am Coll Cardiol, 1992; 19, 267-74.
  • 9Mehmet C, Jim E, Devang P, et al. Long-Term Effects of Novel Biolimus Eluting DEVAXAXXESS Plus Nitinol Self-Expanding Stent in a Porcine Coronary Model. Catheterization and Cardiovascular Interventions, 2006; 68, 271-9.
  • 10Carter A J, Aggarwal M, Kopia GA, et al. Long-term effects of polymer-based, slow-release, sirolimus-eluting stents in a porcine coronary model. Cardiovasc Res, 2004; 63,617-24.

同被引文献5