期刊文献+

静电放电枪校准装置的研制 被引量:4

Development of calibrating apparatus for the ESD simulator
下载PDF
导出
摘要 介绍了自行研制的满足 IEC 61000—4—2标准要求的静电放电模拟器测试系统。在测试系统中使用了极高输入阻抗的高压电压表测量静电放电模拟器的静电高压,提高了静电高压测量准确性;在使用四通道高速取样示波器测量静电脉冲波形中,由于使用了脉冲延时输入与合成迭加的方法,提高了脉冲波形测量的取样率,使测得的脉冲波形更加真实,提高了脉冲波形参数测量的准确性。 A measurement meeting IEC1000-4-2 of testing ESD simulator specification is described. In the system we used a high voltage voltmeter which has very high input impedance to measure electrostatic voltage. So that the measured voltage is accurate. We used simultaneously of digitizing real-time oscilloscopes with four channels to collect the ESD curve. Because methods of pulse delay input and pulse synthesizing are used, the equivalent sampling rate is increased and a more realistic ESD curve is achieved. So we can get accurate parameters of ESD curve.
出处 《宇航计测技术》 CSCD 2000年第1期2-13,共12页 Journal of Astronautic Metrology and Measurement
关键词 测量 脉冲波形 校准 静电放电模拟器 抗静电测试 电磁兼容测试 ESD measurement Simulator Pulse waveform
  • 相关文献

参考文献2

  • 1[1]Basic EMC Publication. Electrostatic discharge immunity test. IEC 1000-4-2 Electromagnetic compatibility(EMC) Part 4: Testing and measurement techniques-Section 2:
  • 2[2]Tektronix. TDS 500C,TDS600B & TDS700C Digitizing oscilloscopes performance verification and Specifications. 070-9874-01

同被引文献37

  • 1张希军,范丽斯,武占成,宋文武.静电模拟器放电开关辐射场特性研究[J].北京理工大学学报,2005,25(z1):68-71. 被引量:10
  • 2周忠元,蒋全兴,张金奎,邓凌翔.静电放电电流波形校准装置的研制与分析[J].东南大学学报(自然科学版),2006,36(3):366-369. 被引量:8
  • 3武占成 刘尚合 等.静电放电模拟器电流波形校验装置研究[J].军械工程学院学报,1999,11(2):1-4.
  • 4Li Huadong.Key factors affecting ESD current waveform[C]// IEEE International Symposium on Electromagnetic Compatibility.New York:The Institute of Electrical and Electronics Engineers Inc,2003:382 -385.
  • 5Centola Federico,Pommerenke David,Wang Kai,et al.ESD excitation model for susceptibility study[C]//IEEE International Symposium on Electromagnetic Compatibility.New York:The Institute of Electrical and Electronics Engineers Inc,2003:58-63.
  • 6International Electrotechnical Commission.IEC 61000-4-2:2001 electromagnetic compatibility (EMC):part 4-2:testing and measurement techniques:electrostatic discharge immunity test[S].Geneva,Switzerland:IEC Central Office,2002.
  • 7American National Standards Institute.ANSI C63.16-1993 American national standard guide for electrostatic discharge test methodologies and criteria for electronic equipment[S].New York:the Institute of Electrical and Electronics Engineers Inc,1994.
  • 8De Ketelaere W,Martens L,Braem Y,et al.Calculation of the ESD-pulse parameters and associated uncertainty for ESD-gun calibration[C]//IEEE International Symposium on Electromagnetic Compatibility.New York:The Institute of Electrical and Electronics Engineers Inc,2000:449-452.
  • 9Agilent technologies.Understanding oscilloscope frequency response and its effect on rise-time accuracy[EB/OL].(2002)[2005-07-10].http://cp.literature.agilent.com/litweb/pdf/5988-8008EN.pdf.
  • 10Sroka Jan.Oscilloscope influence on the calibration uncertainty of the pulse rise time of ESD simulators[C]//IEEE International Symposium on Electromagnetic Compatibility.New York:The Institute of Electrical and Electronics Engineers Inc,2003:378-381.