联合低电压和自动管电流调节技术降低婴幼儿胸部增强CT扫描剂量 被引量:9

Dose reduction for contrast-enhanced chest CT scanning in infants with low tube voltage and automatic tube current modulation technique
下载PDF
导出
摘要 目的评价联合应用低电压和自动管电流调节(ATCM)技术在婴幼儿胸部增强CT扫描中降低辐射剂量、提高血管结构对比的价值。方法对69例3岁以下患儿行64排MSCT增强检查,并分为低电压结合ATCM并噪声指数为9(A组)或10(B组)及标准扫描序列为120kVp结合120mA固定毫安(C组);测量强化主动脉和外周肌肉CT值及噪声,计算SNR和对比噪声比(CNR)。记录容积CT剂量指数(CTDIvol)。采用5分法评估图像质量;以Kappa方法评价观察者间的一致性。结果A组和B组主动脉强化明显,分别较C组增高196.04 HU(65.51%)和186.57 HU(62.34%),A、B组背景噪声平均值较C组分别增加4.15HU和4.36HU;3组间SNR及CNR差异无统计学意义。A组和B组比C组CTDIvol分别降低73.71%和77.00%。结论在婴幼儿64排MSCT增强扫描中联合应用低电压(80kVp)和ATCM技术可获得临床接受的图像质量和良好的对比,并明显降低辐射剂量。 Objective To observe the value of combining low tube voltage and automatic tube current modulation (ATCM) technique for reducing radiation dose and improving contrast vascular structures for contrast-enhanced chest CT scanning in infants. Methods All 69 patients aged 〈3 years were assigned into three groups for contrast-enhanced MSCT scanning, i.e. low kVp and ATCM technique with noise index 9 (group A) or 10 (group B), respectively, and a standard 120 kVp with fixed tube current of 120 mA (group C). Aorta contrast enhancement and surrounding muscle CT values and noises were measured, SNR and contrast-to-noise ratio (CNR) were calculated. The CT dose index of volume (CTDIvol) values were recorded. Imaging quality was assessed by two radiologists using a five-point scale, and the degree of interob-server concordance was evaluated with Kappa statistics. Results Compared with group C, the aorta significantly enhanced in group A and group B, increased by 196.04 HU (65.51%) and 186.57 HU (62. 34%), respectively, and the average background noise increased 4. 15 HU and 4. 36 HU in group A and group B, respectively. SNR and CNR were not statistically different among three groups. Compared with group C, CTDIvol reduced by 73.71% and 77.00% in group A and group B, respectively. Conclusion Combining low tube voltage (80 kVp) with ATCM technique for contrast-enhanced chest CT scanning in infants can obtain good image quality and clinically acceptable contrast degree, and significantly reduce radiation dose.
出处 《中国医学影像技术》 CSCD 北大核心 2014年第5期676-679,共4页 Chinese Journal of Medical Imaging Technology
关键词 儿童 体层摄影术 X线计算机 辐射剂量 自动管电流调节 Child Tomography, X-ray computed Radiation dosage Automatic tube current modulation
  • 相关文献

参考文献16

  • 1Herzog C, Mulvihill DM, Nguyen SA, et al. Pediatric cardiovas- cular CT angiography: Radiation dose reduction using automatic anatomic tube current modulation. AJR Am J Roentgeno|, 2008, 190(5) :1232-1240.
  • 2Feuchtner GM, Jodocy D, Klauser A, et al. Radiation dose re- duction by using 100-kV tube voltage in cardiac 64-slice computed tomography: A comparative study. Eur J Radiol, 2010, 75 (1) : e51-e56.
  • 3Brenner DJ, Hall EJ. Computed tomography--an increasing source of radiation exposure. N Engl J Med, 2007, 357 (22): 2277-2284.
  • 4Kirpalani H, Nahmias C. Radiation risk to children from compu- ted tomography. Pediatrics, 2008,121(2):449-450.
  • 5张祺丰,彭芸,李剑颖,刘玥,曾津津.64排螺旋CT自动管电流调节技术降低儿童胸部扫描剂量的应用研究[J].中国医学影像技术,2009,25(9):1591-1594. 被引量:42
  • 6彭芸,马大庆,李剑颖,张祺丰,刘玥,王蓓,曾津津.64层螺旋CT自动管电流调节技术在婴幼儿胸部低剂量CT扫描中应用的可行性研究[J].中华放射学杂志,2008,42(10):1045-1049. 被引量:40
  • 7Kim JE, Newman B. Evaluation of a radiation dose reduction strategy for pediatric chest CT. AIR Am J Roentgenol, 2010,194 (5) : 1188-1193.
  • 8Bae KT, Tap C, Gurel S, et al. Effect of patient weight and scanning duration on contrast enhancement during pulmonary multidetector CT angiography. Radiology, 2007, 242 (2) : 582- 589.
  • 9Sigal-Cinqualbre AB, Hennequin R, Abada HT, et al. Low-kilo- voltage multi-detector row chest CT in adults: Feasibility and effect on image quality and iodine dose. Radiology, 2004,231 (1) : 169-174.
  • 10Siegel MJ, Schmidt B, Bradley D, et al. Radiation dose and im- age quality in pediatric CT: Effect oftechnical factors and phan- tom size and shape. Radiology, 2004,233(2):515-522.

二级参考文献18

  • 1李剑颖.GE VCT——高图像质量和低剂量的和谐[J].CT理论与应用研究(中英文),2006,15(3):57-60. 被引量:5
  • 2Colang JE, Killion JB, Vano E. Patient dose from CT: a literature review. Radiol Technol, 2007,79 : 17-26.
  • 3Brenner, DJ, Hall EJ. Computed tomography: an increasing source of radiation exposure. N Engl J Med, 2007, 357: 2277 -2284.
  • 4Brody AS, Frush DP, Huda W, et al. Radiation risk to children from computed tomography. Pediatrics ,2007,120:677-682.
  • 5Kalra MK, Rizzo S,Maher MM,et all Chest CT performed with Z-axis modulation : scanning protocol and radiation dose. Radiology, 2005,237 : 303 -308.
  • 6Namasivayam S, Kalra MK, Pottala KM, et al. Optimization of Z-axis automatic exposure control for multideteetor row CT evaluation of neck and comparison with fixed tube current technique for image quality and radiation dose. AJNR,2006,27: 2221-2225.
  • 7Kalra MK, Maher MM, Kamath RS, et al. Sixteen-detector row CT of abdomen and pelvis: study for optimization of Z-axis modulation technique performed in 153 patients. Radiology,2004, 233 : 241-249.
  • 8Kalra MK, Maher MM, D'Souza RV,et al. Detection of urinary tract stones at low-radiation-dose CT with Z-axis automatic tube current modulation: phantom and clinical studies. Radiology, 2005,235:523-529.
  • 9Naidich DP, Marshall CH, Gribbin C, et al. Low-dose CT of the lungs : preliminary observations. Radiology, 1990,175:729-731.
  • 10Itoh S, Ikeda M, Arahata S, et al. Lung cancer screening: minimum tube current required for helical CT. Radiology, 2000, 215:175-183.

共引文献78

同被引文献101

引证文献9

二级引证文献24

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部