期刊文献+

流动注射仪测定水中氰化物两种显色法的比较 被引量:4

Comparison of Two Kinds of Chromogenic Methods for Determining Cyanide in Water by Flow Injection Analyzer
下载PDF
导出
摘要 使用流动注射分析仪分别用吡啶-巴比妥酸和异烟酸-巴比妥酸两种显色方法对水中氰化物进行检测。吡啶-巴比妥酸法检测速度更快,更加适用于大规模检测和水污染事故等应急监测;异烟酸-巴比妥酸法检出限更低,且不需使用吡啶,有利于保护实验人员身心健康。这两种方法均有检测效率高、准确率高、灵敏度高、稳定性高等多种优点。 Flow injection analyzer was used to detect cyanide in water by pyridine-barbituric acid and isonicotinic acid-barbituric acid,the two color methods. The detection rate of pyridine-barbituric acid method was faster,more suitable for mass detection and water pollution accident emergency monitoring. The detection limit of isonicotinic acid-barbituric acid method was lower,without the use of pyridine,which can were protect personnel physical and mental health. These two kinds of methods had advantages of high detection efficiency,high accuracy,high sensitivity and high stability.
作者 毛元宝
出处 《广州化工》 CAS 2014年第9期119-122,共4页 GuangZhou Chemical Industry
关键词 氰化物 流动注射分析 水质检测 cyanide flow injection analysis water quality detection
  • 相关文献

参考文献3

  • 1国家环境保护总局《水和废水监测分析方法》编委会.水和废水监测分析方法:4版[M].北京:中国环境科学出版社,2002:670.
  • 2郭少维,孟祖君,姜鑫,陈杨虹.异烟酸巴比妥酸流动注射法测定水中总氰化物[J].上海师范大学学报(自然科学版),2011,40(6):568-573. 被引量:9
  • 3HJ484-2009,水质氰化物的测定容量法和分光光度法[s].

二级参考文献9

共引文献32

同被引文献22

引证文献4

二级引证文献13