摘要 应用美国哈希COD快速测定仪(进口)与欧陆科仪COD快速测定仪(国产)测定实际水样的氨氮含量。通过与国标(纳氏试剂分光光度法)的实验结果进行对比,通过精密度、准确度及加标回收等实验结果,得出:两种COD快速测定仪测定实际水样的氨氮含量均具有良好的线性关系,虽然灵敏度比国家标准方法低,但精密度和准确度均较高,符合国家检测要求,与国家标准方法的测定结果无显著性差异,而且分析速度快,试剂耗量低,二次污染少,值得在废水监测中推广使用。 The imported fast COD analyzer (HACH, USA) and the domestic COD analyzer (EURO TECH) are used for the determination of ammonia-nitrogen in water samples. Compared with the national standard method, the results of precision, recovery and method showed that the two instruments are reliable with good precision and recovery, and their results are correspondent to the standard method. There are many advantages such as faster analysis speed and lower reagent consumption for the determination of ammonia-nitrogen. So, they can be widely applied to the wastewater monitoring.
出处 《广东化工》 CAS 2014年第9期209-210,共2页 Guangdong Chemical Industry
基金 教育部高等学校博士学科点专项科研基金(Nos.20110171120025) 广东环境保护工程职业学院院长基金资助项目
关键词 氨氮 COD测定仪 快速测定 ammonia-nitrogen COD analyzer rapid determination
  • 相关文献

参考文献12

二级参考文献98

共引文献278

同被引文献33

  • 1余美琼,洪新艺,王碧玉.离子选择电极法测定氨氮浓度[J].福建分析测试,2006,15(3):37-38. 被引量:17
  • 2邓金花,吴清平,廖富迎,王奎兰.环境水质氨氮的快速检测[J].环境监测管理与技术,2007,19(1):33-34. 被引量:18
  • 3HJ535-2009水质氨氮的测定纳氏试剂分光光度法[S].
  • 4HJ168-2010,环境监测分析方法标准制修订技术导则[S].
  • 5GB3838-2002.地表水环境质量标准[S].[S].,..
  • 6GB8978-1996.污水综合排放标准[S].[S].,..
  • 7中华人民共和国环境保护部.水质氨氮的测定水杨酸分光光度法HJ536-2009[S].北京:中华人民共和国环境保护部,2007.
  • 8梁作光,李莹.纳氏光度法现场快速测定环境水中的氨氮[J].化学分析计量,2007,16(6):52-53. 被引量:3
  • 9Asata T, Oikawa K, Kawata K, et al. Ion chromatographic determination of ammonia in air using a sampling tube of porous carbon [ J ]. Analytical Sciences,2004,20(1): 125-128.
  • 10夏玉宇,张完白,李洁,等.化验员实用手册[M].3版.北京化学工业出版社,2012:272-274.