摘要
目的:对胶乳颗粒免疫比浊法测定血清同型半胱氨酸(HCY)的方法学进行初步评价。方法:应用胶乳颗粒免疫比浊法测定HCY的精密度、标准曲线、干扰实验、稳定性实验、并与循环酶法测定血清HCY的结果进行相关性分析。结果:颗粒增强免疫比浊法测定HCY的精密度(批内CV为1.87%、0.75%,批间CV为2.29%、5.35%)较高;HCY浓度在0μmol/L~50μmol/L范围内线性良好;当抗坏血酸浓度1107mg/L、胆红素浓度445 umol/L、血红蛋白浓度6.7 g/L;甘油三酯浓度14.6 mmol/L、氨离子浓度75 umol/L时均未对测定结果产生干扰;试剂开瓶后稳定周期也可达到25天;颗粒增强免疫比浊法和循环酶法测定HCY相关性良好(y =0.9904x +0.0813,r=0.9981)。结论:颗粒增强免疫比浊法测定血清HCY具有较高的精密度,结果准确可靠,适合临床实验室推广。
Objective:To evaluate the detection of serum homocysteine ( HCY) with Particle Enhanced Immunoturbidimetric. Methods:The precision, calibration curve, Anti interference test,on-board stability studies,and the determination of serum HCY and enzymatic cycling assay results by correlation analysis.Results: The Particle Enhanced Immunoturbidimetric method revealed good precision ( CV 1.87%、0.75% , 2.29%、5.35%) .The linear range is0μmol/L~ 50μmol/L.The assay demonstrated less than 10% interference deviation in the presence of Vic-C(1107mg/L), bilirubin(445 umol/L), hemoglobin( 6.7 g/L),TG(14.6 mmol/L), ammoniumion(75 umol/L).samples can be stable after 25 days on-board stability.the Particle enhanced immunoturbidimetric method showed a good correlation with enzymatic cycling assay (y = 0.9904x +0.0813,r= 0.9981).Conclusions:The Particle enhanced immunoturbidimetric method is precise, reliable and suitable for determinations of serum HCY concentration in a clinical laboratory.
出处
《医学检验与临床》
2014年第2期35-37,共3页
Medical Laboratory Science and Clinics
关键词
颗粒增强免疫比浊法
同型半胱氨酸
心血管疾病
方法学评价
Particle Enhanced Immunoturbidimetric
Homocyteine
Cardiovascular disease
Evaluation test of methodology