螺旋断层放疗系统物理性能的验收测试与质量保证 被引量:1

Acceptance Testing and Quality Assurance of Physics Performance in Helical Tomotherapy System
下载PDF
导出
摘要 目的:验收测试新装螺旋断层放疗系统的主要物理性能。方法:参照设备厂家验收手册及AAPM TG148号报告,测试验证螺旋断层放疗系统的剂量学、射束几何学精度及兆伏CT影像测试等物理性能。结果:经测试,多叶光栅T&G测试数值为0.87%;次级准直器Shift/Sweep测试数值为0.05 mm;射束中心轴几何精度偏差数值为-0.377 mm,偏转角度为0.03°;旋转模式下,射束输出稳定性偏差为0.1%;能量测试稳定性偏差为0.8%;固定与移动激光灯精度偏差均小于0.5 mm;多叶光栅中心精度误差为0.86 mm;多叶光栅IEC Y方向平行度偏转角度为0.01°;次级准直器对称性精度偏差值为0.12 mm;射束横向、纵向离轴曲线及百分深度剂量分布与参考数据作γ分析比较,在标准为ΔDD=2%、ΔDTA=1 mm的条件下,γ值均<1;兆伏CT模体吸收剂量<3 cGy,重建影像无伪影,影像分辨率可见密度棒第三排小孔。结论:设备符合各项验收标准,可投入临床使用。 Objective To check the main physics performance of a new helical tomotherapy system. Methods With reference to the acceptance testing procedures of equipment manufacturers and AAPM TG148 report, the acceptance testing was executed for the physics performances of the system in dosiology, geometric accuracy of beam, MVCT image testing and etc. The measurement value of multi-leaf grating T&G was 0.87%, the value of secondary collimator Shift/Sweep was 0.05 mm, the value of geometric accuracy offset of beam axis was -0.377 mm with the angle of 0.03°, the value of beam output stability offset was 0.1% in case of rotating mode, the value of energy test stability offset was 0.8%, the precision offsets of fixed and mobile laser lights were both less than 0.5 mm, the center precision error of multi-leaf grating was 0.86 mm, the angle of parallelism offset along Y direction of multi-leaf grating was 0.01°, and the deviation value of secondary collimator symmetry was 0.12 mm. With the standards of ADD = 2% and △DTA = 1 mm, γ values were all less than 1 when γ analysis was performed for beam off-axis curves along X axis and Y axis, percent depth dose distribution as well reference data. The phantom absorbed dose of MVCT was less than 3 cGy. There were no obvious artifacts in reconstructed image, and the third row of holes in the intensity rod could be distinguished in image resolution. The helical tomotherapy system meets all acceptance standards, and is worth of clinical application.
出处 《医疗卫生装备》 CAS 2014年第5期97-102,共6页 Chinese Medical Equipment Journal
关键词 螺旋断层放射治疗 验收测试 质量保证 影像引导放射治疗 helical tomotherapy acceptance testing quality assurance IGRT
  • 相关文献

参考文献20

  • 1Mac kie T R,Holmes T, Swerd[of S. Tomotherapy : A new concept for the delivery of dynamic conformal radiotherapy[J]. Med Phys, 1993, 20(6):1 709-1 719.
  • 2Sheng K,Molloy J A,Larner J M. A dosimetric comparison of non- coplanar IMRT versus Helical Tomotherapy for nasal cavity and paranasal sinus cancer[J]. Radiother Onco1,2007,82(2) : 174-178.
  • 3崔迪,戴相昆,马林,徐寿平,王运来,周桂霞,冯林春,曲宝林.鼻咽癌螺旋断层放疗与常规加速器调强放疗的剂量学比较[J].中华放射肿瘤学杂志,2008,17(3):169-173. 被引量:29
  • 4Karen Chu, George Rodrigues, Slav Yartsev. Dosimetric evaluation of helical tomotherapy treatment planning for non-small cell lung can- cer[J]. Cancer Therapy, 2008,6 : 571-576.
  • 5Xin Zhang,Penagaricano J,Moros E G. Dosimetric comparison of helical tomotherapy and linac-IMRT treatment plans for head and neck cancer patients[J]. Med Dosimetry, 2009,35 (4) : 264-268.
  • 6李益坤,刘海,王晓萍,杭霞瑜,曾波,甘叶敏,李帆.螺旋断层放疗与常规调强放疗在乳腺癌保乳术后同步推量中的剂量学比较[J].临床肿瘤学杂志,2013,18(7):648-651. 被引量:19
  • 7Kutcher G J, Coia L, Gillin M,et ol. Comprehensive QA for radiation oncology : Report of AAPM Radiation Therapy Committee Task Group 40[J]. Med Phys, 1994,21 (4) :581-618.
  • 8Fenwick J D,Tome W A,Jaradat H A,et al. Quality assurance of a helical tomotherapy machine[J]. Phys Med Biol,2004,49 (13) :2 933- 2 953.
  • 9Balog J,Mackie T R,Pearson D,et ol. Benchmarking beam alignment for a clinical helical tomotherapy device[J]. Med Phys, 2003,30 (6):1 118-1 127.
  • 10Forrest L J, Mackie T R, Ruchala K ,et al. The utility of megavoltage computed tomography images from a helical tomotherapy system for setup verification purposes[J]. Int J Radiat Oncol Biol Phys, 2004,60 (5):1 639-1 644.

二级参考文献22

  • 1洪卫,冉立,卢冰,杨黎,常建英,甘家应,胡银祥.乳腺癌改良根治术后放疗降低患侧肺受量的剂量学研究[J].中华放射医学与防护杂志,2011,31(6). 被引量:17
  • 2戴晓萍,郭建萍.乳腺癌术后放疗病人血清心肌钙蛋白T测定与心脏损伤的研究[J].肿瘤学杂志,2005,11(4):292-293. 被引量:26
  • 3Lee N, Puri DR, Blanco AI, et al. Intensity-Modulated radiation therapy in head and neck cancers: an update. Head Neck,2007, 29:387-400.
  • 4Mackie TR, Holmes T, Swerdloff S, et al. Tomotherapy : a new concept for the delivery of dynamic conformal radiotherapy. Med Phys, 1993,20 : 1709-1719.
  • 5Van Vulpen M, Field C, Raaijmakers CP, et al. Comparing stepand-shoot IMRT with dynamic helical Tomotherapy IMRT plans for head-and-neck cancer. Int J Radiat Oncol Biol Phys, 2005,62: 1535-1539.
  • 6Sheng K, Molloy JA, Read PW. Intensity-modulated radiation therapy(IMRT) dosimetry of the head and neck: a comparison of treatment plans using linear accelerator-based IMRT and helical tomotherapy. Int J Radiat Oncol Biol Phys,2006,65: 917-923.
  • 7Hart C, Liu A, Schultheiss TE, et al. Dosimetric comparisons of helical tomotherapy treatment plans and step-and-shoot intensitymodulated radiosurgery treatment plans in intracranial stereotactic radiosurgery. Int J Radiat Oncol Biol Phys,2006,65: 608-616.
  • 8Sheng K,Molloy JA,Larnera JM,et al. A dosimetric comparison of non-coplanar IMRT versus Helical Tomotherapy for nasal cavity and paranasal sinus cancer. Radiother Oncol,2007,82 : 174-178.
  • 9Wang X,Zhang X,Dong L,et al. Effectiveness of noncoplanar IMRT planning using a parallelized multiresolution beam angle optimization method for paranasal sinus carcinoma. Int J Radiat Oncol Biol Phys ,2005,63:594-601.
  • 10Cozzi L, Fogliata A, Bolsi A, et al. Three-dimensional conformal vs. intensity-modulated radiotherapy in head-and-neck cancer patients: comparative analysis of dosimetric and technical parameters. Int J Radiat Oncol Biol Phys,2004,58:617-624.

共引文献46

同被引文献7

引证文献1

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部