光学相干生物测量仪Lenstar LS 900与IOLMaster对眼球生物学参数测量的一致性评价 被引量:4

Comparison of ocular biometry using Lentar LS 900 and IOL Master
下载PDF
导出
摘要 目的探讨一种最新的光学低相干反射测量仪Lenstar LS900对于眼球生物测量的准确性,将其测量值与光学相干生物测量仪IOLMaster的测量结果进行比较。方法对照观察分别由Lenstar LS900和IOLMaster测量的正常人群眼球的眼轴长度(axial length,AL)、前房深度(anterior chamber depth,ACD)、角膜子午线上下的曲率半径(CR1、CR2)和白对白角膜直径(white—to—white distance,WTW)等眼球生物学参数。结果本研究纳入了206个健康的成年个体(206眼),年龄(26.9±15.6)岁。两种测量仪对于AL、ACD、CR1、CR2,和WTW等测量值的相关程度很高。Lenstar LS900与IOLMaster的测量结果比较,AL差值(0.014±0.090)mm(P=0.025),CR.差值(0.006±0.030)mm(P=0.005),CR,差值(0.020±0.040)mm(P〈0.001),均略长;WTW较短差值(-0.110±0.400)mm(P〈0.001),ACD的差异无统计学意义(P=0.554)。除WTW一致性界限为-0.89~0.67mm外,两种测量仪的测量数据具有良好的一致性。结论与IOLMaster相比较,LenstarLS900可提供同样准确、可靠的眼球生物测量数据(除wTw外),并且可提供中央角膜厚度和晶状体厚度数据,因此LenstarLS900在屈光手术和白内障手术领域有很好的应用前景,能满足未来高质量白内障手术的要求。 Objective To evaluate the ocular biometric measurement using a new optical low-coherence refiectometer ( Leustar LS 900 ), and compare the measurements with those obtained from the IOL Master. Methods In this prospective comparative observational study, axial length ( AL), anterior chamber depth ( ACD ), radii of corneal curvature ( CR1 , CR2 ) and white-to-white distance (WTW) were performed using IOLMaster and Leustar LS 900 biometer by the same examiner. Results The study evaluated 205 healthy adults ( 206 eyes), mean age was ( 25.9 ± 5.6 ) years. There was a good correlation for AL, ACD, CRI, CR2 and WTW between two devices. The Lenstar measured a slightly longer AL, mean difference was ( 0. 014 ±0. 090 )mm ( P = 0. 025 ), mean difference of CR, was ( 0. 005 ±0. 030 ) mm ( P = 0.005 ), mean difference of CR2 was(0.020 ±0. 040 ) nun ( P 〈 0. 001 ), and mean difference of WTW was ( - 0. 110 ±0. 400)mm(P 〈0.001 ). The ACD obtained by the two devices was similar (P = 0. 554). Bland-Altman plots displayed good agreement of all parameters except WTW ( limits of agreement was from -0.89 mm to 0.57 mm). Conclusion The Leustar LS 900 provide precise and valid measurements of all parameters analyzed, which are comparable with those of IOLMaster except WTW. The results, coupled with its availability of central corneal thickness and lens thickness measurements make the Leustar LS 900 a promising new instrument for ocular biometry prior to cataract and refractive surgery.
出处 《眼科新进展》 CAS 北大核心 2014年第6期560-563,共4页 Recent Advances in Ophthalmology
关键词 生物测量 光学低相干反射测量仪 光学相干生物测量仪 biometry optical low-coherence reflectometry partial coherence in-terferometry
  • 相关文献

参考文献16

  • 1Olsen T. Calculation of intraocular lens power: a review[J]. Acta Ophthalmol Scand ,2007,85 (5) :472-485.
  • 2Drexler W, Findl O, Menapace R, Rainer G, Vass C, Hitzenberger CK, et al. Partial coherence interferometry:a novel approach to biometry in cataract surgery [J]. Am J Ophthalmol, 1998,126 (4) :524-534.
  • 3Haigis W, Lege B, Miller N, Schneider B. Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis [J]. Graefes Arch Clin Exp Ophthalmo1,2000 ,238 (9) :755-773.
  • 4Santodomingo-Rubido J, Mallen EA, Gilmartin B, Wofffsohn JS. A new non-contact optical device for ocular biometry[J]. Br J Ophthalmo1,2002,85 (4) :458-452.
  • 5Narvaez J, Cherwek DH, Stulting RD, Waldron R, Zimmerman G J, Wessels IF, et al. Comparing immersion ultrasound with par- tial coherence interferometry for intraocular lens power calculation[J]. Ophthalmic Surg Lasers Imaging,2008,39( 1 ) :30-34.
  • 6Lavanya R, Teo L, Friedman DS, Aung HT, Baskaran M, Gao H, et al. Comparison of anterior chamber depth measurements u- sing the IOLMaster, scanning peripheral anterior chamber depth analyser, and anterior segment optical coherence tomography [J]. Br J Ophthalmol,2007,91 ( 8 ) : 1023-1026.
  • 7Salouti R, Nowroozzadeh MH, Zamani M, Ghoreyshi M, Salouti R. Comparison of the ultrasonographic method with 2 partial coherence interferometry methods for intraocular lens power ca culation [J]. Optometry,2010,82 ( 3 ) : 140-147.
  • 8Olsen T. Prediction of the effective postoperative (intraocular lens) anterior chamber depth [J]. J Cataract Refract Surg ,2006 , 32(3) :419-424.
  • 9Ashwin PT, Shah S, Wolffsohn JS. Advances in cataract surgery [J]. Clin Exp Optom,2009,92(4) :333-342.
  • 10Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement [ J ]. Lancet. 1986.1 (8476) :307-310.

同被引文献45

  • 1刘国军,沈宝莲,王海存,康维强,刘泳,刘志宏,仇宜解.硅油填充眼眼轴的超声测量[J].中华超声影像学杂志,2005,14(7):559-560. 被引量:9
  • 2吴敏,叶剑,孙强,翟军印,曹书,喻箭.前节OCT与A超测量前房深度的比较[J].国际眼科杂志,2007,7(5):1341-1342. 被引量:8
  • 3Rajan MS, Keilhom I, Bell JA. Partial coherence laser interferometry vs conventional ultrasound biometry in intraocular lens power calculations [ J ]. Eye (Lond), 2002, 16(5): 552-556.
  • 4Findl O, Drexler W, Menapace R, et al. Improved prediction of intraocular lens power using partial coherence interferometry [J]. J Cataract Refract Surg,2001, 27(6) : 861-867.
  • 5Bjelos Roncevic M, Busic M, Cima IK,et al. Intraobserver and interobserver repeatability of ocular components measurement in cataract eyes using a new optical low coherence reflectometer [J]. Graefes Arch Clin Exp Ophthalmol, 2010, 249( 1): 83-87.
  • 6Zaldivar R, Shuhz MC, Davidorf JM, et al. Intraocular lens power calculations in patients with extreme myopia[J].J Cataract Refract Surg, 2000, 26(5) : 668-674.
  • 7Tsang CS, Chong GS, Yiu EP, et al. Intraocular lens power calculation formulas in Chinese eyes with high axial myopia [J]. J Cataract Refract Surg, 2003, 29 (7) : 1358-1364.
  • 8Rohrer K, Frueh BE, Walti R, et al. Comparison and evaluation of ocular biometry using a new noncontact optical low-coherence retlectometer[J]. Ophthalmology, 2009, 116( 11 ) : 2087-2092.
  • 9Bjelos Roncevic M, Busic M, Cima I, et al. Comparison of optical low-coherence reflectometry and applanation ultrasound hiometry on intraocular lens power calculation [J]. Graefes Arch Clin Exp Ophthalmol, 2011, 249 ( 1 ) : 69-75.
  • 10Olsen T. Calculation of intraocular lens power: a review [J]. Acta Ophthalmol Seand, 2007, 85 (5) : 472-485.

引证文献4

二级引证文献7

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部