期刊文献+

连续流动分析仪测定水中氰化物影响因素的研究 被引量:2

Influencing Factors on the Determination of Cyanide by Continuous Flow Analyzer
下载PDF
导出
摘要 采用AA3连续流动分析仪分光光度法对环境水中的氰化物进行了测定,且对测定水体中氰化物实验的影响因素进行了研究。实验结果表明,蒸馏温度,实验试剂的配制对实验均有较大的影响,蒸馏温度应选在155℃,吡啶巴比妥酸的配制应注意溶解温度和放置时间的影响。 A continuous flow analyzer was used to determine cyanide in the water. Influencing factors on the determination of cyanide in water were studied. The result indicated that the distillation temperature, the preparation of reagents have a greater impact on the experiment, the distillation temperature should be selected in the 155 ℃, the preparation of pyridine-barbituric acid should choose the appropriate solution temperature and this reagent is not stable.
作者 汪曼
出处 《广东化工》 CAS 2014年第11期221-221,216,共2页 Guangdong Chemical Industry
关键词 AA3连续流动分析仪 氰化物 影响因素 AA3continuous flow analyzer;cyanide;influencing factors
  • 相关文献

参考文献4

二级参考文献7

  • 1.EPA 335.5,吡啶-4-碳酸-1,3-二甲基巴比妥酸分光光度法测定氰化物[S].,..
  • 2.GB 5750-1985.吡啶-巴比妥酸分光光度法测定氰化物[S].,..
  • 3.生活饮用水检验规范[S].[S].,2001..
  • 4.生活饮用水水质卫生规范[S].[S].,2001..
  • 5Standard methods for the examination of water and wastewater. 20th, method-phenols.
  • 6Written by Ninglan Liao Application Group. Quik Chem Method 10- 210 - 00 - 3 - A. 2001.7.17.
  • 7Written by Ninglan Liao Application Group. Quik Chem Method 10 - 204 - 00 - 4 - B. 2001.6.21.