按照PIS-A型瞬态光谱仪的数据格式,用普朗克公式计算1000~6000K黑体的光谱辐射代替实测辐射,分析了减小线性拟合方案测温误差的方法.研究表明,适当选择参与拟合数据波长的上下限、提高信噪比和A/D转换位数,并对维恩近似带来的系统误差进行校正补偿等措施,可有效提高测温精度.
According to the datum form of the PIS-A Instant Spectrometer,the spectral exitance of blackbody of 1000 ̄6000 K was calculated with the Planck's formula to replace the value measured by the PIS-A Instant Spectrometer and some methods of reducing the error of the linear fit scheme were analyzed.This study indicates that a proper selection of upper and lower limits of the wavelengths in fitting operation,the increase of the signal-to-noise ratio and the digits of A/D conversion, and the compensation for the system error caused by the approximation of the Wien's formula to the Planck's formula, can effectively improve the accuracy of measurement.