摘要
目的:采用HPLC法同步测定三七粉中人参皂苷Rg1、Rb1及三七皂苷R1含量。方法:采用SGE protecol C18(5μm,4.6×250mm)色谱柱,流动相为乙腈-水梯度洗脱(0-12min,19∶81;12-60min,19-36∶81-64),检测波长为203nm,流速1.0mL/min,进样量20μL。结果:人参皂苷Rg1、人参皂苷Rb1、三七皂苷R1分别在25.625~430mg/L、26.875~410mg/L、10.625~170mg/L范围呈线性,相关系数r分别为0.9999、0.9998、0.9996;该方法重复性及回收率均符合要求。结论:本法用于同步测定三七粉中人参皂苷Rg1、Rb1及三七皂苷R1的含量,具有简便、准确、高效等特点。
Objective: This study has been to establish a method for synchronous determination of ginsenoside Rg1, ginsenoside Rb1 and notoginsenoside R1 in radix notoginseng powder. Method: SGE protecol C18(5μm,4.6×250mm)chromatographic column was used, with mixtures of acetonitrile and water as mobile phase in gradient mode (0-12min,19∶81; 12-60min, 19-36∶81-64), the wavelength was 203nm, the flow rate was 1.0mL/min, the injection volume was 20μL. Results: The linear range of the three saponins is 25.625 - 430mg/L, 26.875 - 410mg/L, 10.625 - 170mg/L respectively, and the correlation coefficient r is 0.9999, 0.9998, 0.9996 respectively. Repeatability and recovery rate are good. Conclusion: The presented method is a simple, accurate, efficient method for synchronous determination ginsenoside Rg1, ginsenoside Rb1 and notoginsenoside R1 in radix notoginseng powder.
出处
《安徽化工》
CAS
2014年第3期83-84,共2页
Anhui Chemical Industry