摘要
目的:分析并研究瓦里安公司新型直线加速器TrueBeam型号光子束的长期稳定性。方法:采用Standard Imaging公司的QA BeamChecker Plus采集TrueBeam系统6 MV光子线长期日检数据,并进行系统性分析,评价参数包括中心轴剂量、平坦度及对称性。结果:18个月的日检数据分析结果显示,6 MV的X射线中心轴剂量偏差为(0.0%±0.7%),仅1次检测结果>2%;平坦度稳定性好,偏差为(0.1%±0.1%);轴向对称性偏差和横向对称性偏差测量结果均在±1%以内。结论:TrueBeam医用加速器系统的6 MV光子束具有非常稳定的剂量输出特性。
Objective: To study the long-term stability of 6MV photon beam on TrueBeam, which is a new linear accelerator introduced by Varian Medical System. Methods: QA BeamChecker Plus(QABC+, Standard Imaging Inc., Middleton, USA) was used to measure 6MV X-ray on TrueBeam linac as daily quality assurance(Daily QA). Dosimetric parameter, including output Constancy, flatness and symmetry, were employed to evaluate the dose delivery stability of TrueBeam linac. Results: Daily QA results of 18 months showed the dose output delta was(0.0%±0.7%), only once out of tolerance of 2%. The constancy of Flatness was excellent with the delta of(0.1%±0.1%). The symmetric of axial and transverse varied within 1%. Conclusion:The output constancy of TrueBeam with 6MV is very stable, as well as flatness and symmetry. It's stable and reliable to implement Truebeam linear accelerator in clinical.
出处
《中国医学装备》
2014年第6期34-37,共4页
China Medical Equipment
基金
国家自然科学基金面上项目(81171994)"肿瘤转移相关蛋白3(MTA3)对食管-胃交界腺癌淋巴道转移的调控机制"