医用加速器TrueBeam剂量输出长期稳定性研究 被引量:4

Evaluation of dose delivery stability of a new linear accelerator TrueBeam
下载PDF
导出
摘要 目的:分析并研究瓦里安公司新型直线加速器TrueBeam型号光子束的长期稳定性。方法:采用Standard Imaging公司的QA BeamChecker Plus采集TrueBeam系统6 MV光子线长期日检数据,并进行系统性分析,评价参数包括中心轴剂量、平坦度及对称性。结果:18个月的日检数据分析结果显示,6 MV的X射线中心轴剂量偏差为(0.0%±0.7%),仅1次检测结果>2%;平坦度稳定性好,偏差为(0.1%±0.1%);轴向对称性偏差和横向对称性偏差测量结果均在±1%以内。结论:TrueBeam医用加速器系统的6 MV光子束具有非常稳定的剂量输出特性。 Objective: To study the long-term stability of 6MV photon beam on TrueBeam, which is a new linear accelerator introduced by Varian Medical System. Methods: QA BeamChecker Plus(QABC+, Standard Imaging Inc., Middleton, USA) was used to measure 6MV X-ray on TrueBeam linac as daily quality assurance(Daily QA). Dosimetric parameter, including output Constancy, flatness and symmetry, were employed to evaluate the dose delivery stability of TrueBeam linac. Results: Daily QA results of 18 months showed the dose output delta was(0.0%±0.7%), only once out of tolerance of 2%. The constancy of Flatness was excellent with the delta of(0.1%±0.1%). The symmetric of axial and transverse varied within 1%. Conclusion:The output constancy of TrueBeam with 6MV is very stable, as well as flatness and symmetry. It's stable and reliable to implement Truebeam linear accelerator in clinical.
出处 《中国医学装备》 2014年第6期34-37,共4页 China Medical Equipment
基金 国家自然科学基金面上项目(81171994)"肿瘤转移相关蛋白3(MTA3)对食管-胃交界腺癌淋巴道转移的调控机制"
关键词 日检质量控制设备 稳定性 剂量 Daily quality assurance tool Stability Dose
  • 相关文献

参考文献15

  • 1Fourie OL.Comparison of linear accelerator photon outputs from the IAEA TRS-398 and TRS-277 codes of practice[J].Australas Phys Eng Sci Med,2008,31(1):24-31.
  • 2Klein EE,Hanley J,Bayouth J,et al. Task Group 142 report:quality assurance of medical accelerators[J]. Med Phys,2009,36(9) :4197-4212.
  • 3吴丽丽,陆佳扬,陈志坚.鼻咽癌静态调强放疗计划中采用非分裂射野的研究[J].中国肿瘤,2010,19(8):507-510. 被引量:8
  • 4李承军,胡健,王陆州,徐利明.QA BeamChecker Plus的质量保证[J].医疗卫生装备,2010,31(8):120-122. 被引量:4
  • 5Popple RA,Brezovich IA.Dynamic MLC leaf sequencing for integrated linear aocelerator control systems[J]. Med Phys, 2011,38(11) : 6039-6045.
  • 6Zwahlen DR,Lang S,Hrbacek J,et al. The use of photon beams of a flattening filter- free linear accelerator for hypofractionated volumetric modulated arc therapy in localized prostate cancer[J].Int J Radiat Oncol Biol Phys, 2012,83(5): 1655-1660.
  • 7张中柱.中国肿瘤放疗设备的进步——中国医用加速器发展三十年纪[J].中国医学装备,2009,6(2):1-8. 被引量:8
  • 8Hrbacek J,Lang S,Klock S.Commissioning of photon beams of a flattening filter-free linear accelerator and the accuracy of beam modeling using an anisotropic analytical algorithm[J].Int J Radiat Oncol Biol Phys, 2011,80(4): 1228-1237.
  • 9Pai S,Das IJ,Dempsey JF,et al. TG- 69:radiographic film for megavoltage beam dosimetry[J]. Med Phys, 2007,34(6): 2228-2258.
  • 10Alashrah S,Kandaiya S,Lum LS,et al.Depth dependence of the single chamber response function of the I' mRT MatriXX array in a 6MV photon beam[J]. Z Med Phys, 2013,23(4): 270-278.

二级参考文献19

  • 1白彦灵,隋晓红,胡洪涛.放射治疗质量控制应健全运转机制[J].中国肿瘤,2004,13(8):478-479. 被引量:9
  • 2戴建荣,胡逸民,张红志,关莹,张可,王闯.针对患者调强放射治疗计划的剂量学验证[J].中华放射肿瘤学杂志,2004,13(3):229-233. 被引量:65
  • 3肖锋,孙朝阳,石梅,张学成,陆宙,陆军,刘晓莉.鼻咽癌动态调强放疗计划的剂量学验证[J].现代肿瘤医学,2007,15(2):178-180. 被引量:17
  • 4Srivastava SP, Das I J, Kumar A, et al. Dosimetric comparison of split field and fixed jaw techniques for large IMRT target volumes in the head and neck [J]. Med Dosim 2009. (Epub ahead of print).
  • 5Lee CC, WU' A, Garg M, et al. A new approach to reduce number of split fields in large field IMRT[J]. Med Dosim, 2009. (Epub ahead of print).
  • 6Wu" Q, Arnfield M, Tong SD, et al. Dynamic splitting of large intensity-modulate fields [J]. Phy Med Biol 2000, 45 (7):1731-1740.
  • 7Dogan N, Leybovich LB, Sethi A, et al. Automatic feathering of split IMRT fields for step-and-shoot intensity modulated radiation therapy[J]. Phy Med Biol, 2003, 48(9):1133-1140.
  • 8Kamath S, Sahni S, Li J, et al. Generalized field-splitting algorithms for optimal[J]. Phy Med Biol, 2007, 52 (18): 5483-5486.
  • 9Patel L, Natarajom T, Hassan S S .The use of computed radiography for routine linear accelerator and simulator quality control [J]. Br J Radiol, 2009,82 : 827-838.
  • 10ICRU. ICRU report35. Radiation dosimetry:electron beams with energies between 1 and 50MeV[C] // International Commission on Radiation Units and Measurements Bethesda, MD, USA, 1984.

共引文献38

同被引文献23

引证文献4

二级引证文献12

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部