70kV管电压在头颅CT血管成像中的可行性 被引量:8

Feasibility study of cerebral CT angiography using 70kVp tube voltage
下载PDF
导出
摘要 目的:探讨70kV管电压在头颅CT血管成像中的可行性。方法:80例临床疑似脑血管疾病的患者随机分成2组,其中40例患者行新双源CT 70kV头颅CT血管成像(70kV组),另40例患者行120kV头颅CT血管成像(120kV组)。测量每例患者颈内动脉、大脑中动脉CT值,计算信噪比(SNR)和对比噪声比(CNR)。两名放射医师对所有病例以4分法对图像质量进行评分。计算有效剂量(ED)。比较两组间的图像质量及辐射剂量。结果:70kV组颈内动脉及大脑中动脉的平均CT值[(511.1±93.8)HU,(459.1±83.4)HU]高于120kV组的CT值[(289.1±46.6)HU,(260.9±43.4)HU,P<0.001]。70kV组图像噪声[(18.3±2.1)HU]高于120kV组[(8.7±1.0)HU,P<0.001],70kV组的SNR颈内动脉,CNR颈内动脉及SNR大脑中动脉,CNR大脑中动脉(28.3±6.1,26.2±5.9及25.4±5.3,23.3±5.2)低于120kV组(33.9±7.5,30.1±7.1及30.7±7.4,26.9±6.9,P<0.05)。70kV组的主观评分(3.5±0.7)与120kV组(3.6±0.5,P=0.22)无显著性差异;两组读者间一致性为中等(kappa值=0.54,0.59,P<0.001)。70kV组与120kV组ED分别为(0.2±0.0)、(1.2±0.1)mSv(P<0.001),辐射量降低了80%。结论:70kV管电压在头颅CT血管成像中是可行的,能在降低辐射量的同时保持足够的图像质量。 Objective:To explore the radiation dose and image quality of cerebral CT angiography using 70kVp tube voltage.Methods:Eighty patients with clinically suspected cerebrovascular disease were divided into two groups.One group (40 patients)underwent cerebral CT angiography with 70kVp tube voltage,while the other group (40 patients)underwent conventionaltube voltage (120kVp)cerebral CT angiography.The CT values of the internal carotid artery (ICA)and the middle cerebral artery (MCA)in each patient were measured.Signal-to-noise ratio (SNR)and contrast-to-noise ratio (CNR)in each patient were also calculated.Two observers blindly evaluated the image quality using a 4-point scale.Effective dose (ED)was calculated for each patient.Results:The 70kVp group showed higher CT values [(511.1±93.8)HU, (459.1±83.4)HU]than 120kVp group [(289.1±46.6)HU,(260.9±43.4)HU,P〈0.001).Image noise of 70kVp group [(18.3±2.1)HU]was higher than 120kVp group [(8.7±1.0)HU,P〈0.001],the SNRICA,CNRICA and SNRMCA, CNRMCA (28.3±6.1,26.2±5.9 vs 25.4±5.3,23.3±5.2)of 70kV group were lower than 120kV group (33.9±7.5, 30.1±7.1 vs 30.7±7.4,26.9±6.9;P〈0.05).There were no significant differences between the 70kVp group (3.5± 0.7)and the 120kVp group (3.6±0.5,P= 0.22)regarding the mean score.Moderate interreader agreement was found for both groups (kappa values were 0.54 and 0.59,all the P value〈0.001).ED were (0.2±0.0)mSv and (1.2±0.1)mSv for 70kVp and 120kVp groups (P〈0.001)respectively,and the average effective dose was decreased by 80% in the 70kVp group.Conclusion:The application of 70kVp in cerebral CT angiography is feasible,it can reduce radiation dose without noticeable loss in diagnostic quality.
出处 《放射学实践》 2014年第6期585-588,共4页 Radiologic Practice
关键词 血管成像 诊断显像 辐射量 体层摄影术 X线计算机 头部 Angiography Diagnostic imaging Radiation dose Tomography,X-ray computed Head
  • 相关文献

参考文献16

  • 1Brenner DJ, Hall EJ. Computed tomography-an increasing source of radiation exposure[J]. N Engl J Med, 2007,357 (22) : 2277- 2284.
  • 2McCollough CH Primak AN,Braun N,et al. Strategies for reduc- ing radiation dose in CT[J]. Radiol Clin North Am,2009,47(1) : 27-40.
  • 3Zhang WL,Li M,Zhang B,et al. CT angiography of the head and-neck vessels acquired with low tube voltage,low iodine,and itera- tive image reconstruction: clinical evaluation of radiation dose and image quality[J]. PIoS One, 2013,8(12) : 81-86.
  • 4Cho ES,Chung TS, Oh DK, et al. Cerebral computed tomography angiography using a low tube voltage (80kVp) and a moderate concentration of iodine contrast materiaha quantitative and quali- tative comparison with conventional computed tomography angi- ography[J]. Invest Radiol, 2012,47(2) : 142-147.
  • 5Korn A, Bender B, Fenchel M, et al. Sinogram affirmed iterative reconstruction in head CT: improvement of obiective and subjec- tive image quality with concomitant radiation dose reduetion[J]. Eur J Radiol,2013,82(9) :1431-1435.
  • 6Wu TH, Hung SC, Sun JY, et al. How far can the radiation dose be lowered in head CT with iterative reconstruction? Analysis of imaging quality and diagnostic accuracy[J]. Eur Radiol, 2013,23 (9) :2612 -2621.
  • 7Korn A, Fenchel M, Bender B, et al. Iterative reconstruction in head CT:image quality of routine and low-dose protocols in com- parison with standard filtered back-projection[J]. AJNR, 2012, 33(2) 2218-224.
  • 8Moscariello A,Takx RA, Schoepf U J, et al. Coronary CT angiog- raphy:image quality, diagnostic accuracy, and potential for radia- tion dose reduction using a novel iterative image reconstruction technique-comparison with traditional filtered back projection[J]. Eur Radiol,2011,21(10) :2130-2138.
  • 9Prakash P, Kalra MK, Digumarthy SR, et al. Radiation dose reduc- tion with chest computed tomography using adaptive statistical it- erative reconstruction technique: initial experience[J]. J Comput Assist Tomogr, 2010,34 ( 1 ) : 40-45.
  • 10Singh S, Kalra MK, Hsieh J, et al. Abdominal CT : comparison of adaptive statistical iterative and filtered back projection recon- struction techniques[J]. Radiology, 2010,257 (2) : 373-383.

同被引文献65

  • 1ZONDERVAN R L. Frequent body CT scanning of yong adults:indications,outcomes, and risk for radiation - induced cancer [ J ]. J Am CollRadiol, 2011,8 (7) : 501 - 507.
  • 2MAHESH M. Advances in CT technology and application topediatric imaging [J]. Pediatr Radiol,2011,41 (9): 493 -497.
  • 3HUDA W. Technique factors and image quality as funct- ions of patientweight at abdominal CT [J ]. Radiology, 2000,217(2):430-435.
  • 4PFLEDERER T, JAKSTAT J, MARWAN M, et al.Radiation exposureand imagequalityin staged low-dose protocols for coronarydual-sourceCTangiography:arandomized compari- son [ J ].Eur Radiol, 2010,20(5 ): 1 197 - 1 206.
  • 5NAIDICH D P, MARSHALL C H, GRIBBIN C, et al. Low- dose CT of thelungs:preliminary observation [J]. Radiology, 1990,175 ( 3 ) :729- 731.
  • 6SCHINDERA S T,NAUER C,TREIER R, et al. Strategies for reducing the CT radiation dose [J]. Radiologe, 2010,50 (12):1 120-1 127.
  • 7WINKLEHNER A,KARLO C,PUIPPE G,et al.Raw data-based iterative reconstruction inbody CTA:Evaluation of radiation dose saving potential[ J ]. Eur Radiol, 2011,21 (12) :2 512-2 526.
  • 8Leber AW, Becker A, Knez A, et al. Accuracy of 64-slice compu- ted tomography to classify and quantify plaque volumes in the proximal coronary system: a comparative study using intravascular ultrasound[J]. J Am Coll Cardiol,2006,,17(3) :672- 677.
  • 9Herzog C, Mulvihill DM, Nguyen SA, et aI. Pediatric cardiovascu- lar CT angiography: radiation dose reduction using automatic tube current modulation[J]. AJR,2008,190(5) : 1232 -1240.
  • 10Macari M, Bini E J, Xue X,et al. Colorectal neoplasms: prospective comparison of thin section low-dose muhi-detector row CT colonography and conventional colonoscopy for detection[J ]. Ra diology,2002,224 (2) :383-392.

引证文献8

二级引证文献72

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部