低剂量CT扫描结合迭代算法重建行颞骨成像 被引量:4

Radiation dose optimization in temporal bone CT with iterative reconstruction technique
下载PDF
导出
摘要 目的:探讨颞骨低剂量CT扫描结合迭代算法(iDose 3)行颞骨成像的可行性及其最佳参数。方法:以成年尸头为研究对象,应用飞利浦256层CT,采用螺旋扫描模式,分别在电压120kV、100kV和80kV时用不同的管电流时间乘积进行颞骨扫描,从250mAs开始,间隔30mAs逐次降低,直至图像较差不能满足诊断时终止扫描,重建算法采用iDose level 3,滤过函数Y-sharp(C)。两位医生采用5分制评分标准对16个解剖结构进行评估。比较所有图像的图像质量及辐射剂量。结果:"100kV+70mAs+iDose level 3+Y-sharp(C)"的成像组合是良好显示解剖结构的最低剂量组合,有效剂量约0.15mSv。结论:颞骨低剂量扫描切实可行,"100kV+70mAs+iDose level 3+Y-sharp(C)"是最低剂量扫描方案。 Objective:To evaluate the feasibility and the optimal parameters for imaging of the temporal bone by using low-dose-256-slice CT with iterative reconstruction technique(IRT)(iDose 3).Methods:One exsomatized cadaveric head was scanned with spiral scanning mode on 256-slice CT scanner from Philips.Three levels of tube tensions from 120kV to 80kV were chosen for the temporal bone scan.At every tube tension,multiple current-time products were used from 250mAs to 70mAs in 30mAs successive interval decrease.Reconstructions were employed by using IRT (iDose Level 3)and edge-enhangcing reconstruction kenel Y-sharp(C).The visibility of 1 6 anatomical landmarks was scored using a five-point-quality rating.The image quality and radiation dose of every level were assessed.Results:The protocols of"100kV+70mAs+iDose level 3+Y-sharp (C)"can clearly depict the normal anatomy,and the radiation dose is lower than the other else, with effective dose of about 0.15mSv.Conclusion:Low-dose-256-slice CT with IRT is feasible for temporal bone scanning with sufficient anatomical detail.The protocols of"100 kV+70 mAs+iDose level 3+Y-sharp(C)"are the optimized ima-ging protocols.
出处 《放射学实践》 2014年第6期644-646,共3页 Radiologic Practice
基金 上海市科委基础重点继续支持项目(10JC1411500 2010-2013)
关键词 颞骨 辐射剂量 体层摄影术 X线计算机 Temporal bone Radiation dosage Tomography,X-ray computed
  • 相关文献

参考文献7

  • 1Swartz JD,Loevner LA. Imaging of the temporal bone[M]. New York:Thieme, 2009 : 1-6.
  • 2华建明,陈根松,章伟敏.低剂量颞骨多层螺旋CT扫描技术的应用研究[J].中华放射医学与防护杂志,2006,26(5):530-532. 被引量:19
  • 3Niu YT, Mehta D, Zhang ZR, et al. Radiation dose reduction in temporal bone CT with iterative reconstruction technique [J]. AJNR,2012,33(6) :1020-1026.
  • 4Kalra MK, Mather MM, Toth TL. et al. Strategies for CT radia- tion dose optimization[J]. Radiology, 2004,230 (3) : 619-628.
  • 5Siegel M J, Schmidt B, Bradley D, et aL Radiation dose and image quality in pediatric CT:effect of technical factors and phantom size and shape[J]. Radiology,2004,233(2) :515-522.
  • 6杨智云,李树荣,郑少燕,严超贵,彭谦,范淼,陈星,初建平,孟悛非.颞骨64排螺旋CT扫描剂量实验研究[J].临床放射学杂志,2009,28(8):1150-1153. 被引量:13
  • 7Nauer CB,Rieke A,Zubler C,et al. Low-dose temporal bone CT in infants and young children: effective dose and image quality[J]. AJNR,2011,32(8) : 1375-1380.

二级参考文献22

共引文献30

同被引文献28

  • 1华建明,陈根松,章伟敏.低剂量颞骨多层螺旋CT扫描技术的应用研究[J].中华放射医学与防护杂志,2006,26(5):530-532. 被引量:19
  • 2亓恒涛,秦维昌,柳澄,王道才,刘传亚,王巍.64层螺旋CT颞骨低剂量扫描的初步研究[J].中华放射学杂志,2007,41(2):117-120. 被引量:30
  • 3Niu YT, Mehta D, Zhang ZR, et al .Radiation dose reduction in temporal bone CT with iterative reconstruction technique [J].Am J Neuroradiol, 2012, 33(6): 1020 1026.
  • 4Han BK, Grant KLR, Garberich R, et al .Assessment of an iterative reconstruction algorithm (SAFIRE) on image quali- ty in pediatric cardiac CT datasets[J]. J Cardiovasc Comput Tomogr, 2012, 6(3): 200-204.
  • 5Winklehner A. Raw data-based iterative reconstruction in body CTA: evaluation of radiation dose saving potential[J]. Eur Radiol, 2011, 21(12) ;2521-2526.
  • 6Zhao X, Hu J J, Yang T. GPU based iterative cone-beam CT reconstruction using empty space skipping technique[J]. J Xray Sci Technol, 2013, 21(1): 53-69.
  • 7白玫,郑钧正.多排(层)螺旋CT的辐射剂量表达及其影响因素探讨[J].辐射防护,2008,28(1):1-12. 被引量:66
  • 8Y.T. Niu,D. Mehta,Z.R. Zhang,Y.X. Zhang,Y.F. Liu,T.L. Kang,J.F. Xian,Z.C. Wang.??Radiation Dose Reduction in Temporal Bone CT with Iterative Reconstruction Technique(J)American Journal of Neuroradiology . 2012 (6)
  • 9Kalar MK,Mather MM,Toth TL,et al.Strategies for CT radiation dose optimization. Radiology . 2004
  • 10Tan J S P,Tan K-L,Lee J C L,Wan C-M,Leong J-L,Chan L-L.Comparison of eye lens dose on neuroimaging protocols between 16- and 64-section multidetector CT: achieving the lowest possible dose. AJNR. American journal of neuroradiology . 2008

引证文献4

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部