期刊文献+

陀螺仪在全髋置换髋臼杯定位中的应用 被引量:3

Application of gyroscope to acetabular cup positioning in total hip arthroplasty
下载PDF
导出
摘要 背景:全髋置换中良好的髋臼假体方位对于手术的成功起到了至关重要的作用。传统定位工具或精确度欠佳、或太过繁琐,限制了其应用。目的:对比陀螺定位仪与传统定位技术在全髋置换髋臼杯定位中的优劣。方法:应用外侧入路,在同一个塑料髋部模型上由不同经验年限的医生利用传统方法进行了100次髋臼杯定位(每人20次40°外展和15°前倾);再利用陀螺定位仪分别进行了60次相同角度的定位,记录实测角度与事先设定角度的误差。结果与结论:与传统技术相比,使用螺仪定位仪测量的外展角和前倾角误差明显下降。提示陀螺定位仪能使全髋置换中髋臼杯的定位更加精确。 BACKGROUND:Good position of acetabular prosthesis during total hip replacement plays a key role in success of the operation. Traditional location tool has poor accuracy, is too complicated, and limits its application. OBJECTIVE:To compare the advantages and disadvantages of gyroscope with the traditional technique in acetabular component orientation in total hip arthroplasty. METHODS:Using lateral approach, physicians with different experiences used traditional technique to locate acetabular cup 100 times (20 times in each person; 40° abduction and 15° anteversion), and then used gyroscope to locate 60 times at the same angle in the same plastic hip model. The error between the planed and measured values was recorded. RESULTS AND CONCLUSION: Compared to the traditional technique, the error of the abduction and anteversion decreased obviously. These indicated that the gyroscope can make the orientation of acetabular component more precise in total hip arthroplasty.
出处 《中国组织工程研究》 CAS CSCD 2014年第22期3445-3450,共6页 Chinese Journal of Tissue Engineering Research
关键词 植入物 人工假体 全髋置换 假体 前倾 外展 陀螺仪 定位器 髋臼杯 定位 arthroplasty,replacement,hip acetabulum hip prosthesis prosthesis fitting
  • 相关文献

参考文献37

二级参考文献396

共引文献358

同被引文献43

  • 1王海涛.髋臼帽定位器的研制[J].骨与关节损伤杂志,2004,19(8):527-527. 被引量:10
  • 2韩莹莹,肖成双,杨麒巍,来颖,郝爽,马鹤成,李幼琼.CT三维重建下髋臼形态的测量及其意义[J].吉林大学学报(医学版),2011,37(6):1136-1140. 被引量:14
  • 3Bergmann G, Deuretzbacher G, Heller M, et al. Hip contact forces and gait patterns from routine activities [ J ]. J Biomech, 2001,34 (7) :859 -871.
  • 4Lewinnek GE, Lewis JL, Tarr R, et al. Dislocations after total hip-re- placement arthroplasties [ J ]. J Bone Joint Surg( Am), 1978,60 (2) : 217 -220.
  • 5Yoshimine F. The safe-zones for combined cup and neck anteversions that fulfill the essential range of motion and their optimum combina- tion in total hip replacements [ J ]. J Biomech, 2006,39 (7) : 1315 - 1323.
  • 6Gallo J, Havranek V, Zapletalova J. Risk factors for accelerated poly-ethylene wear and osteolysis in ABG I total hip arthroplasty [ J ]. Int Orthop,2010,34(1 ) :19 -26.
  • 7Ulrich SD, Seyler TM, Bennett D, et al. Total hip arthroplasties : what are the reasons for revision[J].Int Orthop, 2008,32 ( 5 ) : 597 - 604.
  • 8Bosco JA, Karkenny AJ, Hutzler LH, et al. Cost burden of 30-day re- admissions following Medicare total hip and knee arthroplasty [ J ]. J Arthroplasty,2014,29 (5) :903 - 905.
  • 9Murlha PE, Hafez MA, Jaramaz B, et al. Variations in acetabular a- natomy with reference to total hip replacement[ J]. J Bone Joint Snrg (Br) .2008.90(3) :308 -313.
  • 10Moskal JT, Capps SG. Improving the accuracy of acetabular compo- nent orientation avoiding malposition [ J ]. J Am Acad Orthop Surg,2010,18 (5) :286 - 296.