摘要
目的建立一种制作简单、成本低廉、性能稳定的小鼠右心室压力检测方法。方法选取一段长度约为15cm的PE-50导管(外径:0.9mm,内径:0.5mm),将一端制成适宜弧度,另一端插入7号注射器针头连接压力换能器。选取80只SPF级雄性C57BL/6小鼠,用自制PE右心室导管行右侧颈外静脉插管至右心室,记录成功例数和每次操作时间。另取40只SPF级雄性C57BL/6小鼠随机分为对照组和慢性低压低氧组,每组20只。慢性低压低氧组小鼠在模拟海拔5000m低压舱内不间断喂养4周,对照组在舱外同时喂养。模型复制成功后,用自制PE右心室导管行右心室插管,记录右心室收缩压;分离左、右心室并称质量,计算Hermann-Willson指数并与右心室收缩压进行相关性分析。结果采用此自制PE右心室导管,右心室插管成功率为90%(72/80),从分离血管至检测出右心室波形所需时间每只3~5rain。慢性低压低氧组右心室收缩压(39.52±4.34)mmHg和Hermann-Willson指数0.356±0.039均显著高于对照组(21.24±2.70)mmHg、0.256土0.020,二者呈显著正相关(P〈0.01)。结论采用本法检测右心室压力,制作简单,使用方便快捷,成功率高,性能稳定且成本低廉,可以推广应用。
Objective To establish a simple,low cost and stable method to detect right ventricular pressure in mice. Methods A PE-50 duct length 15 cm(outside diameter:0.9 ram,inner diameter,.0.5 mm) was bent on one terminal and the other terminal was inserted into a 7 =syringe needle to connect to a pressure transducer. This duct was intubated into right ventricle via right external jugular vein to detect right ventricular pressure in 80 SPF grade male C57BL/6 mice. Successful cases and operation time were recorded. Besides,40 SPF grade male C57BL/6 mice were randomized into the control group(n= 20) and chronic hyperbaric hypoxia group(n= 20). Mice in chronic hyperbaric group were raised in a hyperbaric chamber of simulated 5 000 m high altitude {or 4 weeks. The control group was raised outside the chamber simultaneously. Right ventricular systolic pressure was detected with the PE duct. Left and right ventricles were detached and weighed,and Hermann-Willson index was calculated. Results With this PE duct, right ventricular intubation success rate was 90 % (72/80), the operation cost approximately 3 to 5 min each mouse from the separation of blood vessels to detect the time needed for the right ventricle waveform, right ventricular systolic pressure[(39.52± 4.34 )mm Hg] and Hermann-Willson index(0. 356 ± 0. 039)of chronic hyperbaric hypoxia group were significantly higher than that of control group [-(21.24±2.7)mm Hg and (0. 256±0. 020)] ,which has significant positive correlation (P〈0.01). Conclusion It is simple, fast,stable,costing low and of high success ratio to detect right ventricular pressure with this method.
出处
《重庆医学》
CAS
CSCD
北大核心
2014年第19期2401-2403,共3页
Chongqing medicine
基金
国家自然科学基金资助项目(NSFC81071610)
重庆市自然科学基金资助项目(CSTC2010BB5035)