摘要
本文对电感耦合等离子体原子发射光谱分析 (ICP AES)中电荷注入检测器 (CID)得到的二维光谱的频率特性作了研究。测量光谱经过傅里叶变换得到其光谱的频域分布。结果表明 ,测量光谱在频域中的信号可被分为两部分 :即低频信号和高频信号。前者代表有用的分析信号而后者代表背景和噪声。然而 ,不能采用简单滤除高频信号的方法来去除噪声 ,因为这样虽然高频信号没有明显改变但会使背景变形。
The charactesistics of two-dimension spectra obtained by inductively coupled plasma atomic emission spectrometry (ICP-AES) with charge injection detection (CID) in frequency domain were studied in the present paper. The measurement spectra were Fourier transformed and the frequency distribution of the spectra was obtained. Results showed that the spectra in frequency domain could he divided into two parts:high frequency and low frequency signals. The later stood for measurement spectra and the former for background and noises. However, the high frequecny signals could not be smoothed simply to reduce noises because the background was deteriorated even though the spectral signal did not change significantly.
出处
《光谱学与光谱分析》
SCIE
EI
CAS
CSCD
北大核心
2001年第2期230-232,共3页
Spectroscopy and Spectral Analysis
基金
国家自然科学基金!委员会 (批准号 :2 96750 32 )
教育部优秀年轻教师基金
关键词
电感耦合等离子原子发射光谱分析
电荷注入检测
二维光谱
频域
频率特性
inductively coupled plasma atomic emission spectrometry
charge injection detection
two-dimension spectra
frequency domain