期刊文献+

血浆内毒素定量测定及在感染性疾病中的应用 被引量:4

The Application of Quantitative Examination of Endotoxin in Blood Plasma in the Diagnosis of Infections Diseases
下载PDF
导出
摘要 目的 :初步建立血浆内毒素定量检测模式 ,提出健康人群血浆内毒素正常值及内毒素血症诊断参考值。方法 :用稀释加热法排除血浆干扰 ,以弃G因子鲎试剂在BET 16细菌内毒素测定仪下检测被试者血浆内毒素水平 ,并将结果与临床病原微生物培养比较。结果 :该法强化内毒素回收率 5 0 %~ 15 0 % ,符合美国FDA 1987年版要求。在临床感染 91例患者中 ,内毒素血症 3 7例 ,全部分离出 >1株的G-杆菌 ,其余 5 4例血浆内毒素正常的患者中 ,只有 5例分离出G-杆菌。结论 :该法快速 ,简便 。 Objective: To establish a routine for quanti ta ting endotoxin in blood plasma and determine the reference value of endotoxin in blood plasma of healthy subjects for diagnosis of infectious diseases. Method: The confounding influence of blood plasma was eliminated by dilution and heating. The level of endotoxin in plasma was measured with the BET-16 measuring equipment of bacterial endotoxin by using limulus lysate G fa ctor and compared with to pathogenic microorganism results. Results: The assay can potentiate the endotoxin recovery rate by 50%-150%. S ymboling FDA 1987. of the 91 cases examined, 37 cases were with endotoxemia and more than 1 gram-negative bacteria isolates. The other 54 cases were with norm al level of endotoxin in plasma, of whom only 5 cases were with gram negative ba cteriaisolates. Conclusion: The assay was simple, sensiti ve and quantitative. The method was effective in the determination of gram negat ive bacterial infection and endotoxemia.
机构地区 武昌铁路医院
出处 《医药导报》 CAS 2001年第2期83-85,共3页 Herald of Medicine
基金 铁道部科技基金申报课题
关键词 内毒素 BET-16内毒素测定仪 弃G因子鲎试剂 内毒素血症 感染性疾病 诊断 Andotoxin BET-16 measurementer Exacuanting G factor linulus amedocyte lysate Gram-negative bacterial infection Andotoxe mia
  • 相关文献

参考文献5

  • 1[1]Ada H.The level of endtoxin when the child with lower cinumuty have a tever[J].J Pedist,1986,265-269.
  • 2[2]Sander J H,Harry R P,Jan W T,et al.Endotoxemia: an early predictor of septicaemia in febrile patients[J].Lancet,1988,8586:605-609.
  • 3稻田捷也,铃木美幸,远藤重厚,等.エンドトキシンの血中浓度测定法[J].日本细菌学杂志,1990,45(6):937-942.
  • 4[4]Kinetic L T.Interim guidance for human and veterinary drug products and biologicals[J].Public Health Service,1991.15.
  • 5[5]Robert I R,Francime C L,Jack L.Optimiration of detection of bacterial endotoxin in plasma with the limulus test[J].J Lab Clin Med,1999,116:153-161.

同被引文献36

引证文献4

二级引证文献23