2维和3维采集对脑葡萄糖代谢半定量指标的影响 被引量:2

Effect on Semiquantitive Analysis for Brain Glucose Metabolism with 2 or 3 Dimensional PET Scan Respectively
下载PDF
导出
摘要 目的 研究 2维 (2D)和 3维 (3D)采集对脑葡萄糖代谢半定量指标有无明显影响 ,在实际工作中可否灵活采用 .方法  36例无神经疾病病史、症状和体征 ;脑磁共振成像无异常的受检者分为 2组 ,其中 2D组 15例 ,3D组 2 1例 .静脉注射18F -FDG后行PET脑断层显像 ,经计算机滤波反投影重建获得横断面、冠状面、矢状面断层影像 .横断面影像经处理层厚为 0 .5cm ,按常规分别选取代表双侧额叶、顶叶、颞叶、枕叶、小脑皮质的三层影像 ,勾画出各结构的边界 ,依据体重和注射剂量等参数求出平均SUV值 .将上述各部位的平均SUV值进行左右比较 ,计算出比值 ,称R(L/R) .将各脑叶的平均SUV值与小脑平均SUV值进行比较 ,计算出比值 ,称R(C/Cb) .上述数据进行统计学处理 .结果  2D和3D两组双侧额叶、顶叶、颞叶、枕叶、小脑皮质的平均SUV值 ,R(L/R) 和R(C/Cb) 皆差异无显著性 .结论  2D和 3D采集各有优缺点 ,当采用SUV作为半定量指标对脑葡萄糖代谢状况进行评价时 ,两种采集方式对结果无明显影响 ,故可根据实际情况 。 objective To explore the effect on semiquantitive analysis for brain glucose metabolism with 2 or 3 dimensional PET scan respectively.Methods 36 normal persons were divided into two group,15 for 2D,21 for 3D.PET scans were done after 40 minutes of injection of FDG .The transaxial,the coronal,the saggital were gotten through computer reconstruction. The transaxial slice density was 0.5cm. The borders ofbilateral frontal lobes , bilateral parietal lobes , bilateral tempt lobes ,bilateral occipital lobes and bilateral cerebellum were drawn. According to ROI, weight and injection dose of FDG,mean SUVs in all above lobes were gotten. R(L/R)s from Mean SUVs in the left lobes via those in the right were caculated. R(C/CB)s from mean SUVs in different brain lobes via those different cerebellum lobes were also caculated. Results There was no significance between 2D and 3D about SUV ratios of R(L/R)s and R(C/CB)s . Conculsions The semiquantitive analysis for brain glucose metabolism could not be affected by PET scan methods. According to patients condition , 2D or 3D could be chosen to be used without affecting semiquantitive analysis.
出处 《现代临床医学生物工程学杂志》 2001年第2期85-86,共2页 Journal of Modern Clinical Medical Bioengineering
关键词 2D采集 3D采集 ^18F-脱氧葡萄糖 平均SUV值 PET 脑显像 dimensional(2D) 3-dimensional(3D) 18 F-FDG SUV PET Brain Scan
  • 相关文献

参考文献6

  • 1Strother SC, Casey ME, Hoffman EJ, et al. Measuring PET scannersensitivity: relating countrates to image signal-to-noise ratios using noise equivalentcounts[J].IEEE Trans of Nucl Med,1990,37:783
  • 2Bailey Dl. 3D acquisition and reconstruction in position emission tomography[J].AnnNucl Med,1992,6:123
  • 3Trebossen R,Bendriem B,Bibeiro MJ, et al. Validation of the three-dimensionalacquisition mode in positron emission tomography for the quantitation of [18F]flluoro-DOPA uptake in the human strata[J].J Cereb Blood FlowMetab,1998,18:951
  • 4Vesna Sossi, Terrence R. Quantitative comparison of Three- and Two-Dimensional PETwith human brain studies[J].J Nucl Med,1998,39:1714
  • 5Orihara H,Ishii K,Iwata R,et al.Three-dimensional position emission tomographyimaging system and its application for medical use[J].Nippon Rinsho,1997,55:2148
  • 6Aylman RR,Kison PV,Whal RL,et al. Capabilities of two and three-dimensionalFDG-PETfor detecting small lesions and lymph nodes in the upper torso:a dynamic phantomstudy[J].Eur J Nucl Med,1999,26:39

同被引文献22

  • 1王骏,赵海恩.影响PET图像质量的几个相关因素[J].放射学实践,2006,21(2):197-199. 被引量:1
  • 2李天然,高荣光,钱根年,郑春雨,李铭山,王楷堂.PET-CT在神经系统疾病诊断中的应用[J].临床军医杂志,2006,34(1):99-102. 被引量:5
  • 3王欣璐,王全师,吴湖炳,王巧愚,李洪生.应用全身^(18)F-FDG PET/PET-CT为脑转移瘤患者寻找原发灶[J].实用放射学杂志,2006,22(9):1130-1133. 被引量:7
  • 4Turkington T G. lnlroduction to PET instrumentation[J]. J Nucl Med Technol,2001,29(1) :4 11.
  • 5Czernin J, Schelbert H. PET/CT imaging: facts, opinions, hopes and questions[J], J Nucl Med, 2004, 45(Suppl 1): 1S-3S.
  • 6Townsend D W, Carney J P. Yap J T, et ah PET/CT today and tomorrow[J]. J Nucl Med, 2004, 45(Suppl 1):4 S- 14S.
  • 7Wagner H N. Highlights 2002 lecture: reinventing clinical nuclearmedicine[J]. J Nucl Med, 2002, 43(8): 11N-39N.
  • 8Sachelarie I, Kerr K, Ghesani M, et al. Integrated PET-CT: evidence-based review of oncology indications[J]. Oncology,2005, 19(4):481-490.
  • 9Khan N, Oriuchi N, Higuchi T, et al. PET in the follow-up of differentiated thyroid cancer [J]. Br J Radiol, 2003, 76 (910) :690-695.
  • 10Strother S C, Casey M E, Hoffman E J. Measuring PET scanner sensitivity: relating countrates to imagesignal-tonoise ratios using noise equivalents counts[J]. Nucl Sci, 1990,37(2):783-788.

引证文献2

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部