摘要 目的 :比较不同方法测量LASIK术前后角膜厚度的差异。方法 :分别用JEMED/EGH 40 0 0型超声角膜测厚仪和Orbscan Ⅱ眼前节系统于LASIK术前及术后一月测量角膜中心厚度各三次 ,取其均值比较。结果 :LASIK术前两种方法测量结果差异无显著性意义 (P >0 .0 5 )。术后Orbscan Ⅱ眼前节分析诊断系统测量中心角膜厚度值低于超声角膜测厚仪 ,两者比较差异有显著性意义 (P <0 .0 5 )。结论 :LASIK术后 ,角膜弯曲度、球面性的改变及其设计理论 ,可能影响角膜地形图的测定结果。对于欠矫病例再手术时 ,应用两种方法测量角膜厚度 ,互相参照 ,为再次手术提供保障。 Objective:To compare corneal thickenss value before and after LASIK operation with two different measurement.Methods:JEMED/EGH 4000 ultrosonic corneal thickness measurement and Orbscan Ⅱ anterior system were used to measure the central corneal thickness one month before and after LASIK.Results:There was no significant difference between the two methods before LASIK(P>0.05),but they had significant difference(P<0.05) after LASIK, with Orbscan Ⅱ anterior system show lower values than those of ultrasonic meaurement in central corneal thickness. Conclusion:After LASIK, corneal topography may be affected by corneal sphere,curvature change and the design. A comparative analysis of the two values can offer reference for the undercorrected cases when they were designed to be re operated.
作者 刘磊 李新宇
出处 《眼视光学杂志》 2001年第1期8-9,共2页 Chinese Journal of Optometry & Ophthalmology
关键词 角膜地形图 角膜测厚仪 角膜厚度 LASIK comeal topography instrument of corneal thickness central corneal thickness
  • 相关文献

参考文献1

二级参考文献4

共引文献13

同被引文献25

  • 1陈家祺,王铮,杨斌,李绍珍.准分子激光原位角膜磨镶术治疗近视[J].中华眼科杂志,1998,34(2):141-145. 被引量:100
  • 2李风鸣主编.眼科全书.第八卷.第1版[M].北京:人民卫生出版社,1996.2519—2521.
  • 3Nader G,Iskander,Ellen Anderson Penno.Accuracy of Orbscan pachymetry measurements and DHG ultrasound pachymetry in primary laser in situ keratomileusis and LASIK enhancement procedures[J].J Cataract Refract Surg,2001,27(5):681-685.
  • 4Liu ZG,Andrew J,Pflugfelder SC,et al.Evaluation of cornealthickness and topography in normal eyes using the Orbscan cornealtopography system[J].Br J Ophthalmol,1999,83(8)∶774-778.
  • 5Yaylali V,Kaufman SC,Thompson HW.Corneal thickness measurements with the Orbscan topography system and ultrasonic pachymetry[J].J Cataract Refract Surg,1997,23(7):1345-1350.
  • 6Nissen J,Hjortdal J,Ehlers N,et al.A clinical comparisonof optical and ultrasonic pachometry[J].Acta Ophthalmol(Copenh),1991,69(4):659-663.
  • 7Seiler T,Koufala K,Gichter G.Iatrogenic keratectasiaafter laser in situ keratomileusis[J].J Refract Surg,1998,14(4):312-317.
  • 8Joo CK,Kim TG.Corneal ectasia detected after laser insitu keratomileusis for correction of less than 12 diopters of myopia[J].J Cataract Refract Surg,2000,26(2):292-295.
  • 9[1]Liu ZG, Andrew J, Pflugfelder SC, et al. Evaluation of corneal thickness and topography in normal eyes using the Orbscan corneal topography system. Br J Ophthalmol, 1999,83:774
  • 10[4]Douthwaite WA, Matilla MT. The TMS-1 corneal topography measurement applied to calibrated ellipsoidal convex surfaces. Cornea, 1996,15:147

引证文献6

二级引证文献20

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部