CT剂量指数的研究 被引量:7

Study of the Computed Tomography Dose Index.
下载PDF
导出
摘要 目的 研究用不同积分区间 (± 7T和± 5 0mm)计算CT剂量指数 (CTDI)时对结果的影响。方法 用热释光剂量计 (TLD)在标准的CT剂量监测模体上 ,测量日立W -10 0 0型CT机的单层扫描的剂量分布曲线 ,根据分布曲线计算不同积分区所对应的CTDI值。结果 对层厚大于 7mm(10mm)的单层扫描 ,± 5 0mm积分长度所得的CTDI10 0 小于± 7T积分长度所得的CTDIF;对层厚小于 7mm(5mm)的单层扫描 ,则CTDI10 0 大于CTDIF。结论 当扫描层厚不是 7mm时 ,用 10 0mm活性长度的笔形电离室测得的CTDI10 0 与CTDIF 有差异 ,为了便于统一比较 。 Objective The impact on the computed tomography dose index(CTDI)value of varying integrating lengths(±7 T and±50 mm)is analyged.Methods A Hitachi W-1000 scanner was used with thermoluminescent dosimeters and computed tomography dosimetry phantoms.The dose profiles are measured for single scan and the corresponding CTDI calculated for varying integrating lengths.Results For slice thickness over 7 mm(10 mm),the CTDI 100 value obtained with ±50 mm of integrating length will underestimate the CTDI F.By contrast,for the slice thickness below 7 mm(5 mm),the CTDI 100 will overestimate the CTDI F.Conclusion The differences between CTDI 100 value and CTDI F are presented for varying slice thickness.The measured CTDI 100 value should be corrected for different slice thickness when using a pencil ionization chamber.
出处 《中国辐射卫生》 2001年第2期65-66,共2页 Chinese Journal of Radiological Health
基金 香港城市大学科研基金资助课题
关键词 CT剂量监测模体 热释光剂量计 CT剂量指数 层厚 Computed tomography dosimetry phantom Thermoluminescent dosimeter Computed tomography dose index Slice thickness
  • 相关文献

参考文献5

  • 1[1]Code of Federal Regulations[ R]. 21 CFR 1020,33. Health and Human Services Bulletin,41-1- 93 Edition.
  • 2[2]Leitz W,Axelsson B,Szendro G. Computed tomography dose assessment-a practical approach [ J ].Radiation Protection Dosimetry, 1995,57: 377-380.
  • 3[3]Conway BJ,McCrohan JL,Antonsen RG,et al. Average radiation dose in standard CT examinations of the head: results of the1990 NEXT survey[J]. Radiology 1992,184:135-140.
  • 4[4]Knox HH,Gagne RM. Alternative methods of obtaining the computed tomography dose index[ J ]. Health Phys, 1996,71 (2) :219-224.
  • 5[5]IAEA,ILO,WHO, et al, International Basic Standard of Radiation Protection and The Safety of Radiation Sources[ S]. IAEA Safety. Series No, 115 ,Vienna, 1996.

同被引文献45

引证文献7

二级引证文献60

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部