期刊文献+

溶胶凝胶溴离子选择电极的研制与应用 被引量:5

Development and Applications of Bromine Ion Selective Electrode Based on the Sol-Gel Technique
下载PDF
导出
摘要 报道了一种以溶胶凝胶为载体的溴离子选择电极。电极信号响应线性范围为1 .0× 1 0 - 1~ 1 .0× 1 0 - 5mol/L,斜率为 5 7m V/pc;检出限为 2 .5× 1 0 - 6 mol/L。该电极响应快 ,体积小 ,稳定性和重现性好。电极用作测定溴离子浓度 。 The sol gel technique has a great future for immobilization. This paper reports the sol gel as a carrier for bromine ion selective electrode. The electrode showed Nernst response for bromine ion in the linear range of 1.0×10 -1 ~1.0×10 -5 mol/L with a slope of 57 mV/p c . The detection limit is 2.5×10 -6 mol/L. The electrode has the advantages of fast response, small size, good stability and repreducibility. The electrode was used to determine the bromine ion concentration. The results are satisfactory.
出处 《分析科学学报》 CAS CSCD 北大核心 2001年第4期297-299,共3页 Journal of Analytical Science
基金 陕西省教育厅自然科学基金 (No. 0 0 JK1 2 8)
关键词 溶胶凝胶 离子选择电极 溴离子 测定 制备 Sol gel Ion selective electrode Bromine