PET图像定量计算中的误差分析 被引量:3

Analysis of Noise on PET Images
下载PDF
导出
摘要 目的 :分析PET影像中的误差成分 ,以期为分析PET定量计算精度提供思路。材料与方法 :PET ,质控模型 ,1 8F FDG ,在 5个剂量用 2D方式采集数据。计算1 8F FDG均匀分布区域像素计数均值和标准差 ,曲线拟合。结果 :综合误差平方随信号变化呈二次曲线型。误差成分可分为三部分 :( 1)与随机符合相关的成分 :正比于信号强度 ;( 2 )与放射性统计涨落相关的成分 :正比于信号强度平方根 ;( 3)其它成分 :与信号强度无关。结论 :和γ照相机和SPECT图像中的信噪比随信号源强度一直增加不同 。 Purpose: In order to properly conduct quantitative measurement with PET, components of noise were studied. Materials and Methods: ECAT EXACT HR +PET (Siemens) scanner and a quality control phantom (Toshiba) was used. Image data was collected in 2D mode at five doses of 18 F FDG. Pixel mean counts and standard deviation were calculated within ROI in the uniformity area of the phantom. Curve fitting was done with the data. Results: The square of synthetic noise on a PET image varies as signal and the relationship could be expressed with quadrative polynomial;  2=0 0395 N 2+0 1427 N+0 0025 (R 2=0 9358). The synthetic noise may be categorized into three components. The first one varies directly with signal intensity and is related to random coincidence; the second one was related to the square root of signal intensity and was related to Poisson fluctuation; and the third one does not relate to signal intensity. Conclusion: The noise arise mainly from the second component related to Poisson fluctuation; but with the increase of signal intensity, the proportion of the second component decreases and the proportion of the first component related to random conicidence increases.
出处 《中国医学影像学杂志》 CSCD 2001年第5期352-354,共3页 Chinese Journal of Medical Imaging
关键词 PET 图像 误差 定量计算 PET image noise
  • 相关文献

参考文献1

共引文献4

同被引文献23

  • 1姜继伟,章斌,桑士标.SPECT/CT装机后早期质控的必要性探讨[J].中国血液流变学杂志,2014(1):172-173. 被引量:3
  • 2耿建华,陈英茂,陈盛祖,田嘉禾,何义杰,乔士珍.符合线路显像与PET显像中SUV的比较研究[J].中华核医学杂志,2004,24(5):308-309. 被引量:11
  • 3万良荣,黄钢,孙晓光,刘建军,施一平.PET/CT的质量控制和日常维护[J].上海第二医科大学学报,2005,25(12):1303-1305. 被引量:12
  • 4Budinger TF. PET instrumentation: what are the limits. Semin Nucl Med, 1998,28(3) :247-267.
  • 5Zhang H, Inoue T, Tian M, et al. A basic study on lesion detectability for hot spot imaging of positron emitters with dedicated PET and positron coincidence gamma camera. Ann Nucl Med, 2001,15(3) : 301-306.
  • 6Zaidi H. Comparative evaluation of scatter correction techniques in 3D positron emission tomography. Eur J Nucl Med, 2000, 27:1813-1826.
  • 7Bergstrom M, Eriksson L, Bohm C, et al. Correction for scattered radiation in a ring detector positron camera by integral transformation of the projections. J Comput Assist Tomogr, 1983, 7:42-50.
  • 8Bailey DL, Meikle SR. A convolution-subtraction scatter correction method for 3D PET. Phys Med Biol, 1994, 39: 411-424.
  • 9Levin CS, Dahlbom M, Hoffman EJ. A Monte Carlo correction for the effect of Compton scattering in 3D PET brain imaging. IEEE Trans Nucl Sci, 1995,42:1181-1188.
  • 10Hutton BF, Baccarne V. Efficient scatter modelling for incorporation in maximum likelihood reconstruction. Eur J Nucl Med, 1998, 25:1658-1665.

引证文献3

二级引证文献11

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部