合肥市医用诊断X射线机质量控制检测与评价

Test and Evaluation of Quality Control of Medical Diagnostic X-ray Equipment in Hefei
原文传递
导出
摘要 为调查本市医用诊断X射线机运行状况。我站使用RMP型医用诊断X射线诊断影像质量控制检测箱及FJ307G1直读式X、γ照射量仪,检测本市各级医院的101台透视和摄片X射线机的14项影像质量控制检测指标。结果在14项检测指标中有7项指标合格率低于70%。各项检测指标都合格的机器仅13台,其它机器均有一项或几项指标不合格。作者认为目前本市医用诊断X射线机运行状况不良。 Objective: To examine working state of medical diagnostic X-ray equipment in Hefei . Method: 14 quality control targets of 101 X-ray equipments from hospitals at various levels in Hefei were tested with RMP medical diagnostic X-ray image quality control examination kit and FJ-307G 1 instrument which could read X, γ-ray irradiation dose directly. Result: The results indicated that qualify rate of 7/14 targets is lower than 70%. Only 13 equipments are qualified according to the quality control criterion. Conclusion: It is urgent and necessary to carry out quality control of medical diagnostic X-ray equipment in order to improve their working state.
出处 《安徽预防医学杂志》 1998年第2期129-132,共4页 Anhui Journal of Preventive Medicine
关键词 X射线机 质量控制 质量检测 放射卫生 合肥市 medical diagnostic X-ray quality control