摘要
以 1,2 二苯乙烯 4,4′ 双 (1 偶氮 ) 3,4 二羟基苯 2 ,2′ 二磺酸铵 (芪唑 )为柱前衍生试剂 ,采用反相离子对高效液相色谱法分离和测定钪、锡和铝。在C18柱上 ,用含 2 0mmol/LHAc NaAc缓冲溶液 (pH 6 0 )和 10mmol/L十二烷基磺酸钠的甲醇水溶液作流动相 ,检测波长 5 0 0nm ,同时分离测定了钪、锡和铝的络合物。钪、锡和铝的检出限分别为 0 9μg/L ,1 0 μg/L和 1 2 μg/L ,此方法用于矿石样品分析获得满意结果。
The reversed phase ion pair high performance liquid chromatographic determination of scandium, tin and aluminum with diammonium stilbene 4,4′ bis(1 azo) 3,4 dihydroxybenzene 2,2′ disulfate as precolumn chelating reagent on Spherex C 18 column has been investigated with the detection wavelength of 500 nm. Various parameters such as pH, reagent concentration, ion pair reagents, interference of foreign ions have been studied. The mobile phase was methanol water (40∶60, V/V ) containing 20 mmol/L acetate buffer of pH 6 0 and 10 mmol/L sodium dodecyl sulfonate. Three complexes could be completely separated within 17 min. The method has been successfully used for the determination of the three metals in mineral samples ,with recoveries of 98%, 101%, 103% respectively, and RSD was less than 3 6%. Beer's Law was obeyed over the concentration range 0 01 mg/L 3 80 mg/L,0 06 mg/L 4 50 mg/L and 0 04 mg/L 3 40 mg/L respectively. The detection limits were 0 9 μg/L, 1 0 μg/L and 1 2 μg/L for scandium, tin and aluminum when the ratio of signal to noise was 3 Most of other metals did not interfere with the determination The present method is sensitive and selective
出处
《色谱》
CAS
CSCD
北大核心
2001年第5期472-474,共3页
Chinese Journal of Chromatography
基金