摘要
对Packard 2000CA型液体闪烁仪器,“效率示踪DPM”校正猝灭的效果不如外标准转换谱指数法(tSIE)好。对于严重化学猝灭和颜色猝灭样品,“效率示踪DPM”分别给出较标准值偏高和偏低的校正值。
The results of quenching correction by efficiency tracing DPM are worse than by tSIE in Packard 2000CA liquid scintillation counter. Efficiency tracing DPM gives greater and smaller than standard values for strong chemical and colour quenching samples respectively
出处
《核电子学与探测技术》
CAS
CSCD
北大核心
1989年第4期193-196,共4页
Nuclear Electronics & Detection Technology
关键词
液体闪烁计数
猝灭校正
效率示踪
Liquid scintillation counting, Quenching correction, Efficiency tracing DPM