摘要 目的 评定JUSTY II型全自动根尖电测仪在根管工作长度测定中的准确性。方法 选择 80例病牙 ,12 0个根管 ,采用全自动根尖电测仪对根管工作长度进行测定 ,并与插针X线摄片法进行对比。结果 根尖电测仪测定根管工作长度的准确率为 89.17%,插针X线摄片法测定根管工作长度的准确率为34 .17%,二组有显著性差异 (P <0 .0 1)。结论 应用全自动根尖电测仪测定根管工作长度具有准确性高 ,减少患者X线损害 。 Objective To Assess the accuracy of determining root canal length by Justy II full auto root apex locator made by Toei Electric Co. Ltd. Japan.Methods 80 cases and 120 root canals were selected, root canal lengths were determined by apex locator and comparing accuracy with reamer x ray. Results accuracy rate of determining root canal length by apex locator was 89.17% and with reamer x ray was 34.17%. There was significant difference between the two groups ( P <0.01).Conclusion determining root canal length by full auto apex locator could attain good accuracy, reduce x ray harm to patients, increase success rate of root canal treatment and decrease clinical time.
出处 《口腔材料器械杂志》 2001年第4期197-199,共3页 Chinese Journal of Dental Materials and Devices
关键词 根尖电测仪 根管工作长度 准确性 口腔医学 Apex locator Length of root canal Accuracy
  • 相关文献

参考文献4

  • 1黄力子.根管长度电测法及其原理[J].中华口腔医学杂志,1988,23(5):297-297.
  • 2周原能,许佩瑜,邹映青.电测法和手感法测量根管长度准确性的比较[J].现代口腔医学杂志,1991,5(2):123-124. 被引量:24
  • 3张世国 张玉坛 等.利用阻抗测量技术测量牙齿根管长度的研究[J].现代口腔医学杂志,1988,2(4):200-202.
  • 4赵皿 谢欣梅 等.拔牙实测根管长度100例分析测算根管工作长度电测法的准确性[J].实用口腔医学杂志,1987,3(1):30-31.

二级参考文献6

  • 1许佩瑜,周原能.拔牙实测法观察牙科发声探针仪的准确性[J].临床口腔医学杂志,1989,5(1):14-15. 被引量:1
  • 2智能化多功能口腔阻抗仪问世[J]现代口腔医学杂志,1988(04).
  • 3张世国,张玉增,杨彩文,黄力子.利用阻抗测量技术测量牙齿根管长度的研究[J]现代口腔医学杂志,1988(04).
  • 4张世国,张玉增,史亚军,杨彩文,王作哲,刘亚男,黄力子.KZC-3型口腔阻抗测定仪的研制[J]现代口腔医学杂志,1988(02).
  • 5赵皿,谢欣梅,黄力子.拔牙实测根管长度100例分析兼评根管长度电测法的准确性[J]实用口腔医学杂志,1987(01).
  • 6黄力子.关于根管长度电测法原理的实验研究[J]华西口腔医学杂志,1983(02).