摘要
目的 :探讨螺纹异体皮质骨椎间融合支架 (allograftinterbodyfusionCageAIFC)腰椎间融合的稳定性与强度。方法 :用 4具新鲜尸体的尺骨、桡骨和腓骨制成AIFC ,同时用环钻取带双面皮质的髂骨行压缩试验 ,并模拟临床切除L2~ 3 ,L4~ 5椎间盘后分别将AIFC和髂骨植入椎间隙行拔出试验。结果 :( 1)最大抗压负荷AIFC组为 ( 82 6 8± 490 3)N ,髂骨组为( 1385± 2 2 3)N ,AIFC组为髂骨组的 6倍 (P <0 .0 1)。 ( 2 )植骨块移位时的最大拔出力AIFC组为 ( 4 89± 2 0 6 )N ,髂骨组 ( 89± 45 )N ,AIFC组是髂骨组的 5 .5倍 (P <0 .0 1)。结论 :AIFC具有足够的支撑、维持或增加椎间隙高度功能 ,椎间稳定性好 ,其设计独特合理 。
Objective: To measure strength and intervertebral stability of using allograft interbody fusion cage(AIFC)for lumbar interbody fusion.Methods:The AIFC were made of ulnar,radius,fibula from 4 individual donors and cylindrical iliac bone were harvested simultaneously with the trephine cutter from iliac crests,the compression tests of the material were performed in vitro.Resection of the L 2~3 and L 4~5 intervertebral discs were done with the trephine cutter to simulate a lumbar surgical procedures,AIFC and iliac bone blocks were inserted respectively in the intervertebral spaces and segment pull out tests were performed.Results:Maximum compressive load and drawing load were 8268±4905N and 498±206N respectively in AIFC group,but these were 1385±223N and 89±45N respectively in the iliac bone group,it was significant greater in AIFC group than in iliac bone group( P <0.01).Conclusion:AIFC could provide enough support,anti slide ability and could remain or increase the height of intervertebral space.The design is distinctive and reasonable.It completely meets biomechanical and clinical requirements.
出处
《中国矫形外科杂志》
CAS
CSCD
2001年第10期990-992,共3页
Orthopedic Journal of China
基金
兰州军区资助课题 (LXH99- 1 9)
关键词
腰椎
同种
异体植骨
椎间融合
生物力学
Lumbar vertebrae
Homologous
Allograft
Interbody fusion
Biomechanics