摘要
研究了反相高效液相色谱法 (RP HPLC)直接测定茶叶水提取物、可乐饮料中的咖啡因的新方法 ,采用PE C1 8柱 ,以CH3OH H2 O HAc( 30 :69 92 :0 0 8)溶液作流动相 ,紫外检测波长为 2 60nm。可在 8min内将咖啡因与基体分离 ,咖啡因含量与峰面积在 1 0 μg/mL~ 1 0 0 μg/mL范围呈线性关系 ,回归方程为 :Y =335 67c( μg/mL) - 1 2 642 1 ,r=0 .9997。该法可用于可乐饮料。
A new method for determination of caffeine in tea and cola by reversed phase high performance liquid chromatographic(RP HPLC) with MeOH H 2O HAc as the mobile phase was proposed. A good seperation of cafeine with matrix was obtained within 8 min by using PE C18 column. The proposed method was applied to monitor and analyze the content of caffeine in teas and cola drinks with satisfactory results.
出处
《分析试验室》
CAS
CSCD
北大核心
2002年第1期68-70,共3页