摘要 目的 观察 1~ 4个衔铁引起MRI伪影的程度。方法 实验在 0 .15Tesla永磁型磁共振机上进行 ,样本选用JJ 型磁性固位体的衔铁部分 ,在志愿者口腔内进行矢状面扫描 ,对MRI成像进行测量并分析其最大径。结果  1~ 4个衔铁引起伪影的最大径在 10 1~ 10 9.6mm范围内。结论 即使在患者口腔内同时安装 4个衔铁 ,进行MRI检查时产生的伪影仍十分局限。 Objective This research was designed to observe the degree of artifacts caused by 1-4 keepers.Methods Samples were imaged in a living human subject by 0.15 Tesla MR devices in sagital plane. Results were analyzed and the maximal diameters of artifacts were measured.Results The maximal diameters of the artifacts caused by 1-4 keepers were 101-109.6mm.Conclusion Even with four keepers in the dental arch, the degree of artifact is still limited.
出处 《上海口腔医学》 CAS CSCD 2001年第4期316-318,共3页 Shanghai Journal of Stomatology
关键词 衔铁 磁性固位体 磁共振成像 伪影 义齿 Keeper Magnetic Attachment MRI Artifact
  • 相关文献

参考文献5

  • 1[1]limuro FT. Magnetic resonance imaging artifacts and the magnetic attachment system[J]. Dent Mater J,1994,13(1):76-88.
  • 2[2]Laurell KA, Gegauff AG, Rosenstiel SF, et al. Magnetic resonance images degradation from prosthetic magnet keepers[J].Prosthetic Dentistry,1989, 62:344-348.
  • 3魏斌,张富强,余强.磁性固位体引起MRI伪影的研究[J].口腔颌面修复学杂志,2000,1(4):200-202. 被引量:8
  • 4[4]Wingerchuk DM, Krecke KN, Fulgham JR. Multifocal brain MRI artifacts secondary to embolic metal fragments[J]. Neurology,1997, 49(5):1451-1453.
  • 5魏斌.磁性固位体与核磁共振的相互影响[J].国外医学(口腔医学分册),2000,27(3):164-165. 被引量:2

二级参考文献2

  • 1Laurell K.A,GEGAUFF A.G,Rosenstiel S.F,et al.Magnetic resonance images degradation from prosthetic magnet keepers[].The Journal of ProsThetic Dentistry.1989
  • 2赵铱民,欧阳官,高元,金同春.新型磁性固位体的研制[J].实用口腔医学杂志,1991,7(2):70-73. 被引量:9

共引文献8

同被引文献35

  • 1吴国锋,赵铱民,王宝成,沈丽娟,越野.两种铸造温度对铸接式衔铁性能的影响[J].中华口腔医学杂志,2001,36(6):37-39. 被引量:7
  • 2魏斌,钱琼,张富强.磁性附着体义齿衔铁的设计[J].上海第二医科大学学报,2004,24(12):1042-1044. 被引量:2
  • 3孙樱林,刘玉华.镍铬合金固定修复体对MRI影像的影响[J].现代口腔医学杂志,2007,21(1):72-74. 被引量:17
  • 4Svetlize CA,Bodereau EF Jr.Comparative study of retentive anchor systems for overdentures[J].Quintessence Int,2004,35(6):443-448.
  • 5Kokubo Y,Fukushima S.Magnetic attachment for esthetic management of an overdenture[J].J Prosthet Dent,2002,88(3):354-355.
  • 6Watanabe I,Tanaka Y,Ohkubo C,et al.Application of cast magnetic attachments to sectional complete dentures for a patient with microstomia:a clinical report[J].J Prosthet Dent,2002,88(6):573-577.
  • 7Tokuhisa M,Matsushita M,Koyano K.In vitro study of a mandibular implant overdenture retained with ball,magnet,or bar attachments:Comparison of load transfer and denture stability[J].Int J Prosthodont,2003,16(2):128-134.
  • 8Ohkubo C,Watanabe I,Tanaka Y,et al.Application of cast iron-platinum keeper to a collapsible denture for a patient with constricted oral opening:A clinical report[J].J Prosthet dent2003,90(1):6-9.
  • 9Ishikawa M,Kashiwabara T,Ishida O,et al.Installing magnetic keepers using LASER welding[J].J Prosthodont,2002,11(1):49-52.
  • 10Xu C,Chao YL,Du L,et al.Measurements of the flux densities of static magnetic fields generated by two types of dental magnetic attachments and their retentive forces[J].Sichuan Da Xue Xue Bao Yi Xue Ban,2004,35(3):412-415.