摘要
目的测量点扫描质子束流水中深度剂量分布,检测最大剂量点与穿透深度的重复性。方法利用PEAKFINDER测量低能、中能和高能质子束流的水中深度剂量分布,给出深度分布曲线。结果最大剂量点与穿透深度值重复性的最大标准差分别为0. 05、0. 03 mm,远低于允许值1 mm。结论 PEAKFINDER测的最大剂量点与穿透深度精确度和重复性满足检测要求。
Objective To measure the depth dose distribution of proton under spot scanning in water,testing the reproducibility of the depth of maximum dose value and depth of penetration. Methods Measure the depth dose distribution of proton in low、middle and high energy scope with the PEAKFINDER,get the depth distribution curve. Results The maximum standard deviation of repeatability of maximum dose value and penetration depth is 0. 05 mm and 0. 03 mm,respectively,which is lower than the permitted value 1 mm. Conclusion The reproducibility and precision of the depth of maximum dose value and depth of penetration fulfill the testing request of PEAKFINDER.
作者
李明生
袁继龙
姜庆寰
程金生
LI Mingsheng;YUAN Jilong;JIANG Qinghuan;CHENG Jinsheng(National institute for radiological protection,China CDC,Bering 100088 China)
出处
《中国辐射卫生》
2018年第5期483-485,489,共4页
Chinese Journal of Radiological Health
关键词
质子
点扫描
深度剂量分布
质量控制检测
Proton
Spot Scanning
Depth Dose Distribution
Quality Control Testing