湖北省7台医用直线加速器调强放射治疗多叶光栅叶片到位精度验证方法研究 被引量:2

Verification of positioning accuracy of MLC leaves of 7 medical linear accelerators in IMRT in Hubei province
原文传递
导出
摘要 目的用放射性免冲洗胶片测量调强放射治疗(IMRT)多叶光栅(MLC)叶片到位精确度验证方法研究,为IMRT质量控制提供参考依据。方法选择湖北省7家三甲医院的7台不同型号医用直线加速器,放射治疗计划系统(TPS)计划5条宽度为0.6cm条状的栅栏,每条状野之间的距离为3.0cm。用EBT2免冲洗胶片测量5条状野位置、多叶光栅叶片位置偏差和实际宽度。结果按国际原子能机构(IAEA)要求,胶片测量与TPS计划每条栅栏野多叶光栅条状位置结果应≤±0.5mm。编号5和编号7的加速器MLC条状野位置偏差分别为0.7和-1.0mm,不符合IAEA要求。按IAEA要求,胶片测量每对与每条所有多叶光栅叶片平均位置偏差应≤±0.5mm。7台加速器MLC每对叶片位置偏差均在±0.5mm内,符合IAEA要求;IAEA要求单对MLC叶片开野宽度与该机器5条条状野平均开野宽度的差值不超过±0.75mm,7台加速器每对MLC开野宽度与平均值的差值范围为-0.6~0.5mm,符合要求。所有叶片开野宽度标准偏差应≤0.3mm,7台加速器标准偏差在0.1~0.2mm范围内,符合要求。结论用EBT2免冲洗胶片验证MLC位置精确度方法操作简单、检测精度高,是质量控制(QA)的重要手段之一,适合大范围核查使用。 Objective To verify the leaf positioning accuracy of multi-leaf collimator (MLC) in intensity modulated radiation therapy (IMRT) by using radiochromic films to provide references for IMRT quality control. Methods Medical linear accelerators of different designs, owned by 7 first-class hospitals at grade-3 in Hubei, were selected for this verification study. Five strip picket fence pattern was created by treatment planning system (TPS). Each strip was 0.6 cm wide with a 3 cm strip separation. MLC leaf positions, positioning bias and opening widths were then measured with EBT2 radiochromic films. Results According to IAEA standards, the film-measured and TPS planned MLC leaf position difference was ±0.5 mm. The difference of MCL strip position of No.5 and No. 7 accelerator was 0.7 and -1.0 mm, respectively, not in line with the IAEA standards. The difference of film-measured MLC leaf position between each pairs and all pairs of leaves of 7 accelerators were all within ± 0.5 mm, in line with the IAEA requirements. As required by IAEA, the difference of opening width of MLC leave of each pair relative to the averaged widths of all pairs should be within ±0.75 mm. The filem-measured values from 7 accelerators ranged from -0.6 to 0.5 mm, all in line with the IAEA standards. Standard deviation of opening width of all leaves were required to be within 0.3 mm. The measured values from 7 accelerators ranged from 0.1 to 0.2 mm, in line with the IAEA standard. Conclusions The use of EBT2 radiochromic films to verify MLC leaf positioning accuracy is an important means of quality control owing to its simplicity and high measurement accuracy. It is recommended for verification use at a large scale.
作者 周文珊 喻洁 孙刚涛 叶松 马新兴 Zhou Wenshan;Yu Jie;Sun Gangtao;Ye Song;Ma Xinxing(Department of Radiological Health Monitoring and Evaluation,Hubei Provincial Center for Disease Control and Prevention,Wuhan 430079,China)
出处 《中华放射医学与防护杂志》 CAS CSCD 北大核心 2019年第2期132-136,共5页 Chinese Journal of Radiological Medicine and Protection
基金 IAEA资助项目(17821/CRP).
关键词 调强放射治疗 多叶光栅叶片 放射性免冲洗胶片 到位精确度 Intensity modulated radiation therapy MLC leaves Radiochromic film Positioning accuracy
  • 相关文献

参考文献3

二级参考文献33

  • 1Lossasso T, Chui CS, Ling CC. Comprehensive quality assurance for the delivery of intensity modulated radiotherapy with a multileaf collimator used in the dynamic mode. Med Phys, 2001, 28:2209-2219.
  • 2Essers M, de Langen M, Dirkx MLP, et al. Commissioning of a commercially available system for intensity-modulated radiotherapy dose delivery with dynamic multileaf collimation. Radiother Oncol, 2001,60:215-224.
  • 3Budgell GJ, Mort JH, Williams PC, et al. Requirements for leaf position accuracy for dynamic multileaf collimation. Phys Med Biol, 2000, 45:1211-1227.
  • 4Saw CB, Ayyangar KM, Thompson RB, et al. Commissioning and quality assurace for MLC-based IMRT. Med Dosim, 2001, 26:125-133.
  • 5United Nations Scientific Committee on the Effects of Atomic Radiation. Sources and effects of ionizing radiation, report to the General Assembly, Annex B [ R ]. New York: United Nations, 2000 : 84-156.
  • 6World Health Organization. WHO handbook on indoor radon: a public health perspective[ R]. Geneva: WHO, 2009.
  • 7Nazaroff WW, Doyle SM, Nero AV, et al. Potable water as a source of airborne 222Rn in U. S. dwellings: a review and assessment[ J]. Health Phys, 1987, 52 (3) :281-295.
  • 8United Nations Scientific Committee on the Effects of Atomic Radiation. Effects of ionizing radiation, report to the General Assembly, Annex E [ R ]. New York: United Nations, 2006 : 201-334.
  • 9KOrner S, Trautmannsheimer M, Htibel K. Investigation and reduction of personnel exposure levels in Bavarian water supply facilities [ C ]. International Congress Series, 2005, 1276, 240 -241.
  • 10Ringer W, Simader M, Bernreiter M, et al. Mitigation of three water supplies with high radon exposure to the employees [ J ]. Radiat Prot Dosim, 2008, 130(1 ) : 26-29.

共引文献13

同被引文献11

引证文献2

二级引证文献7

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部