期刊文献+

不同材料组配式带膜血管内支架的比较研究 被引量:4

Comparison of various configurations of dacron and politef-covered stents:application to modular endovascular systems
原文传递
导出
摘要 目的 通过比较涤纶 (D)、聚四氟乙烯 (PTFE)及可膨胀性聚四氟乙烯 (ePTFE)覆盖内支架不同组配的漏出量及结合点的阻力 ,确定带膜血管内支架的最佳材料。方法 研究分 5组 ,分别为涤纶组 1 (D1 )、PTFE组 1 (PTFE1 )、涤纶组 2 (D2 )、PTFE组 2 (PTFE2 )及可扩张PTFE组(ePTFE) ,每组含有 5个样本。每一样本由 1个PalmazP2 94内支架和内置、覆盖两段D、PTFE及ePTFE组成。各组以不同的方式进行组配。增加装配内压力 [1 0 0、2 0 0及 30 0mmHg(1mmHg=0 .1 33kPa) ]来验证其外漏 ,在装配两端安置拉力验证其结合点的阻力。方差分析 (ANOVA)用于组间比较。结果 在增加装配内压力时有 4个组配崩溃 (D1 3个 ,D2 1个 )。每 1 5s的漏出量在D1为 (38.4± 2 2 .3)ml,D2为 (40 .4± 1 0 .3)ml;PTFE1、PTFE2及ePTFE无漏出。在牵拉实验中各组在如下拉力下崩溃 ,D1 (4.86± 2 .66)N ,D2 (4.76± 2 .36)N ,PTFE1 (2 0 .49± 2 .80 )N ,PTFE2 (2 2 .0 3± 1 .88)N ,ePTFE (1 8.85± 2 .62 )N。结论 PTFE是带膜血管内支架的较理想材料 。 Objective To determine which graft material is the better choice for endovascular stent graft by comparing the leakage and resistance to traction of stents covered internally and externally with either Dacron (D),polytetrafluoroethylene (PTFE),or expanded PTFE (ePTFE).Methods Individual Palmaz P294 stents were assembled with collagen impregnated Dacron,PTFE or ePTFE covering the stent in one of 5 configurations (D1,D2,PTFE1,PTFE2,ePTFE).D1 and PTFE1 had a 7 mm diameter graft covering two thirds of the stent internally and another cover externally along the entire length of the stent.The same diameter grafts were used in D2 and PTFE2,in which the grafts covered the entire stent internally and externally;The ePTFE grafts were also totally covered the stent,but with 4 mm diameter graft that had been dilated to 8?mm.Resistance to leakage was tested by increasing the internal pressure (100,200 and 300?mm?Hg) with water injections;Traction was placed at both ends of the assembly to test resistance of the junction.Group comparisons were performed using ANOVA.Results Five samples of each configuration were tested.During the pressure test,4 samples collapsed (3 in D1 and 1 in D2).The amount of leakage per 15 s was (38.4±22.3)?ml for D1 and (40.4±10.3)?ml for D2 respectively;There was no leakage measured for PTFE1,PTFE2 and ePTFE.In the traction test,the samples collapsed under a force of (4.86±2.66) Newtons (N) in D1,(4.76±2.36)?N in D2,(20.49± 2.80)?N in PTFE1,(22.03±1.88)?N in PTFE2,(18.85±2.62)?N in ePTFE respectively.There was significant difference between Dacron and the PTFE groups in leakage and resistance.Conclusion PTFE stent PTFE configurations had no leakage under pressure and high resistance to traction as compared with stents covered with Dacron,suggesting that PTFE is a better material for an endovascular stent graft.ePTFE had the same characteristics as normal PTFE,but is easier to assemble in vivo.
出处 《中华实验外科杂志》 CAS CSCD 北大核心 2002年第3期220-222,共3页 Chinese Journal of Experimental Surgery
关键词 涤纶 聚四氟乙烯 组配式带膜血管内支架 比较研究 Dacron Polytetrafluoroethylene Endovascular stent graft Modulity
  • 相关文献

参考文献10

  • 1Parodi JC,Palmaz JC,Barone HD.Transfemoral intraluminal graft implantation for abdominal aortic aneurysm.Ann Vasc Surg,1991,5:491-499.
  • 2Blum U,Voshage G,Lammer J,et al.Endoluminalstent-grafts for infrarenal abdominal aortic anurysms.N Engl J Med,1997,336:13-20.
  • 3May J,White GH,Yu W,et al.Concurrent comparison of endoluminal versus open repair in the treatment of abdominal aortic aneurysms:Analysis of 303 patients by life table method.J Vasc Surg,1998,27:213-221.
  • 4White GH,May J,McGahan T,et al.Historic control comparison of outcome for matched groups of patients undergoing endoluminal versus open repair of abdominal aortic aneurysms.J Vasc Surg,1996,23:201-212.
  • 5Brewster DC,Geller SC,Kaufman JA,et al.Initial experience with endovascular aneurysm repair:comparison of early results with outcome of conventional open repair.J Vasc Surg,1998,27:992-1005.
  • 6Harris EJ.Modular system in the treatment of abdominal aortic aneurysms:Lessons learned in the development of designer endografts.Semin Vasc Surg,1999,12:170-175.
  • 7Malina M,Brunkwall J,Lindblad B,et al.Endovascular management of the juxtarenal aortic aneurysm:can uncovered stents safely cross the renal arteries? Semin Vasc Surg,1999,12:182-191.
  • 8Chuter TAM,Reilly LM.Surgical reconstruction of the iliac arteries prior to endovascular aortic aneurysm repair.J Endovasc Surg,1997,4:307-311.
  • 9Umscheld T,stelter WJ.Time-related alterations in shape,position,and structure of self-expanding,modular aortic stent-grafts:a 4-year single follow-up.J Endovasc Surg,1999,6:17-32.
  • 10Malina M,Lindblad B,Ivancev K,et al.Why do aortic stent-grafts migrate? Experimental aspects.Presented at the 16th annual meeting of the association of international vascular surgeons.Meribel France,1998,3:22-28.

同被引文献59

  • 1秦黎明,马小川,李延静.64层CT成像在心血管疾病中的临床应用[J].包头医学院学报,2006,22(3):349-352. 被引量:4
  • 2乔树宾,侯青,徐波,陈珏,刘海波,杨跃进,吴永健,袁晋青,吴元,戴军,尤士杰,马卫华,张沛,高展,窦克非,邱洪,慕朝伟,陈纪林,高润霖.紫杉醇支架和金属裸支架治疗弥漫长病变的效果比较研究[J].中国循环杂志,2006,21(5):333-336. 被引量:10
  • 3张志辉,李伟杰,王海昌,吕安林,贾国良,李成祥,秦涛,王晓燕.地塞米松涂层支架临床应用的观察[J].心脏杂志,2007,19(2):192-193. 被引量:7
  • 4范德增,贾志远,严心浩,刘晓蕾,董伟,孙福玉,计剑,徐建平,任科峰,陈伟东,沈家骢,邱红,高润霖.类细胞膜仿生药物缓释涂层冠脉支架材料的研究[J].生物医学工程学杂志,2007,24(3):599-602. 被引量:3
  • 5Beck T, Burgstahler C, Reimann A, et al. Technology insight: possible applications of multislice computed tomography in clinical cardiology. Nat Clin Pract Cardiovasc Med. 2005;2(7):361-368.
  • 6Hamoir XL, Flohr T, Hamoir V, et al. Coronary arteries: assessment of image quality and optimal reconstruction window in retrospective ECG-gated multislice CT at 375-ms gantry rotation time. Eur Radiol. 2005;15(2):296-304.
  • 7Foley JB, Brown RI, Penn IM. Restenosis following elective implantation of single Palmaz-Schatz stents in de-novo lesions in native vessels. Coron Artery Dis. 1994;5(1):73-80.
  • 8Bolz A, Schaldach M, Schricker KT. In vitro studies of the blood compatibility of amorphous silicon carbide. Biomed Tech (Berl). 1990;35 Suppl 3:25-27.
  • 9Gutensohn K, Beythien C, Bau J, et al. In vitro analyses of diamond-like carbon coated stents. Reduction of metal ion release platelet activation, and thrombogenicity. Thromb Res. 2000;99(6):577-585.
  • 10Meireles GC, de Abreu LM, Forte AA, et al. Randomized comparative study of diamond-like carbon coated stainless steel stent versus uncoated stent implantation in patients with coronary artery disease. Arcl Bras Cardiol. 2007;88(4):390-395.

引证文献4

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部