摘要 从信背比、干扰等效浓度、干扰校正因子几方面讨论了铁对多种元素的干扰 ,提出了干扰校正方法 ,用于实际样品测定 ,结果满意。 The spectral interferences of iron in the determination of various elements by axial view ICP AES, including signal to background ratio, interference equivalent concentration (IEC) and interference correction factor, were discussed thoroughly in this paper. A method of interference correction was proposed and applied satisfactorily to the analysis of ferric chloride which was used as a water treatment agent. RSD′s found were within the range of 0.96%~4.10% ( n =10), and the recoveries obtained were in the range of 89.5%~118.0%.
出处 《理化检验(化学分册)》 CAS CSCD 北大核心 2002年第5期240-242,共3页 Physical Testing and Chemical Analysis(Part B:Chemical Analysis)
关键词 校正 ICP-AES 三氯化铁 净水剂 光谱干扰 分析 电感耦合等离子体发射光谱 ICP AES Spectral interference Iron Water treatment agent (FeCl 3)